RIGHT is a randomized, prospective, two-arm study that will assess the differential efficacy of VITALITY 2 implantable cardioverter defibrillators (ICDs) using Rhythm ID™ versus selected Medtronic ICDs.
RIGHT will compare the incidence of Ventricular Tachycardia/Fibrillation (VT/VF) therapies delivered for Supraventricular Tachycardia (SVT) episodes using VITALITY 2 devices with Rhythm ID™ to that of ICDs employing other rhythm discrimination algorithms. Appropriateness of therapies will be assessed after independent adjudication of all VT/VF events leading to therapy. Comparisons will be made between device manufacturers across patient populations receiving either a dual- or single-chamber device.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
PREVENTION
Masking
NONE
Enrollment
1,962
VITALITY 2 ICD
Maximo, Marquis, Intrinsic, Virtuoso, or Entrust ICD
Medical University of South Carolina
Charleston, South Carolina, United States
Number of Patients With Inappropriate Ventricular Tachycardia (VT)/Ventricular Fibrillation (VF) Therapy (Shock or Antitachycardia Pacing [ATP]) After the Pre-Discharge Visit
An inappropriate therapy is defined as a VT/VF therapy, either shock or antitachycardia pacing (ATP), delivered for a supraventricular tachycardia (SVT). All events for which a VT/VF therapy was delivered and a stored electrogram exists were reviewed by an independent adjudication committee to determine the appropriateness of device rhythm classification and subsequent therapy delivery.
Time frame: From date of pre-discharge until a minimum of 12 months follow-up until study closure
Time to First Inappropriate Shock Using VITALITY and Selected Medtronic Implantable Cardioverter Defibrillator (ICDs)
Time frame: Time of event
Positive Predictive Value (PPV) of Ventricular Tachycardia/Fibrillation (VT/VF) Discrimination Algorithms Found in VITALITY 2 and Medtronic Implantable Cardioverter Defibrillators (ICDs)
Time frame: Time of event
Time to First Inappropriate Therapy for Which the Discrimination Algorithm Found in VITALITY 2 and Medtronic Implantable Cardioverter Defibrillator (ICDs) Inappropriately Classified the Episode
Time frame: Time of event