The purpose of this study is to evaluate the effect of individual adjustment of the postventricular atrial blanking period in avoiding inappropriate mode switch of dual chamber pacemakers.
Far-field R-wave sensing (FFS) in the atrial channel of dual chamber pacemakers is a relevant source for inappropriate mode switch from the DDD mode to the DDI or VDI mode. Inappropriate loss of atrioventricular synchrony due to false positive mode switch is hemodynamically disadvantageous, may induce atrial tachyarrhythmias, can lead to pacemaker syndrome, and impairs the reliability of pacemaker Holter data. The aim of the study is to determine whether individual adjustment of the postventricular atrial blanking period (PVAB) based on an additional test is effective in avoiding inappropriate mode switch due to FFS when compared to standard programming of the PVAB
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
PREVENTION
Masking
SINGLE
Enrollment
207
Performance of a test to detect far-field R-wave sensing
nominal pacemaker settings
Deutsches Herzzentrum Muenchen
Munich, Germany
Incidence of inappropriate mode switch due to far field R-wave sensing
Time frame: 3 months
Incidence of inappropriate mode switch due to far field R-wave sensing in relation to the lead position
Time frame: 3 months