* A disposable blood pressure measuring device will be tested and compared to a standard mercury sphygmomanometer. * Up to 100 subjects with various levels of blood pressure values will be tested.
The study will comprise 85 to 100 subjects with blood pressure values ranging between 90 and 170mmHg -systolic and 65-95 -diastolic. * Three sets of measurements will be performed on each subject. The measurement with the tested device placed between twh determinations with a mercury sphygmomanometer. * Each measurement will be performed simultaneously by two examiners specially trained. * Informed consent will be obtained from all participants. * The results will be evaluated by a standard statistical plan ANSI/AAMI SP-10 2002.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
SINGLE
Enrollment
120
Department of Medicine, Maaynei Hayesha Hospital
Tel Aviv, Israel
RECRUITINGThe difference between the paired measurements with the tested device and the reference measurements (with a mercury sphygmomanometer) will differ by no more than +-5mmHg.