The main objective of this study is to assess the safety and effectiveness of the sirolimus-coated Bx VELOCITY™ stent in maintaining minimum lumen diameter in de novo native coronary artery lesions as compared to the uncoated Bx VELOCITY™ balloon-expandable stent. Both stents are mounted on the Raptor® Stent Delivery Systems. The secondary objective is to assess cost-effectiveness expressed in incremental cost/life year gained or cost/quality adjusted life year gained at different time points (8 months, 1 year, 3 and 5 years).
This is a multicenter ,prospective, randomized double blind study. This study has a 2-arm design assessing the safety and effectiveness of the sirolimus-coated Bx VELOCITY™ stent to the uncoated Bx VELOCITY™ stent, both mounted on Raptor® Stent Delivery Systems. A total of 100 patients will be entered in the study and will be randomized on a 1:1 basis. Patients who meet the eligibility criteria will be either randomized to Treatment A or Treatment B. Neither the investigator nor the patient will know which stent will be implanted. Patients will be followed at 30 days, 6, 9, and 12 months, and at 2, 3, 4 and 5 years post-procedure, with all patients undergoing repeat angiography at 8 months. Additionally, medical costs associated with the index hospitalization and length of stay, and repeat hospitalizations and costs associated with other relevant medical resource use during the 5 years follow-up period will be collected and analyzed.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
DOUBLE
Enrollment
100
sirolimus-coated Bx VELOCITY Balloon-Expandable Stent
un-coated Bx VELOCITY Stent
in-stent minimum lumen diameter (MLD)
Time frame: 8 months
composite of Major Adverse Cardiac Events defined as death, myocardial infarction (Q wave and non-Q wave), emergent bypass surgery, or repeat target lesion revascularization
Time frame: 30 days and 6, 9, and 12 months, and 2, 3, 4 and 5 years
angiographic binary restenosis (³50% diameter stenosis)
Time frame: 8 months
in-lesion MLD
Time frame: 8 months
target lesion revascularization
Time frame: 9 months
target vessel revascularization
Time frame: 9 months
target vessel failure defined as cardiac death, myocardial infarction, or target vessel revascularization
Time frame: 9 months
procedure success defined as achievement of a final diameter stenosis of <50% (by QCA) using any percutaneous method, without the occurrence of death, MI, or repeat revascularization of the target lesion
Time frame: up to hospital discharge
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.