The aim of this study was to determine the effectiveness of endoscopic full-thickness plication (Plicator; NDO Surgical, Inc., Mansfield, MA) for the treatment of GERD in comparison to a sham procedure. Patients with symptomatic GERD requiring maintenance proton pump inhibitor therapy were entered into a randomized, single-blind, prospective, multicenter trial. Seventy-eight patients were randomly assigned to undergo endoscopic full-thickness restructuring of the gastric cardia with transmural suture. Eighty-one patients underwent a sham procedure. Group assignments were revealed following the 3-month evaluation. The primary endpoint was ≥ 50% improvement in GERD-HRQL score. Secondary endpoints included medication use and esophageal acid exposure. Patients achieving ≥ 50% improvement in GERD-HRQL score at 3-months versus baseline off-meds were considered responsive to their assigned procedure. Patients who failed to reach this level of improvement at 3-months were considered non-responders. Analysis of these dichotomized variables (responder/non-responder) was done using Fisher's exact test comparing the proportion of responders between the active and sham groups. Intent-to-treat analysis was also performed. The null hypothesis was that the proportion of responders was the same in both the active and sham groups. Testing was done at the 5% level of significance (alpha=0.05).
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
DOUBLE
Enrollment
159
A low-profile (≤6mm) gastroscope was advanced and an EGD followed by Savary wire placement was performed. The gastroscope was then removed and the Plicator was advanced directly over the wire and into the stomach. The gastroscope was re-introduced through the Plicator, and the Plicator was retroflexed under direct endoscopic visualization to the anterior gastric cardia, approximately 1cm below the gastroesophageal (GE) junction. The Plicator arms were opened and the endoscopic tissue retractor was advanced deeply into the gastric cardia. The gastric wall was then retracted into the open arms of the Plicator. The arms were closed and a single, transmural pledgeted suture was deployed. The Plicator and gastroscope were then removed and the gastroscope re-inserted to evaluate the post-plication GE junction.
The sham procedure was identical to the treatment procedure through the positioning of the retroflexed Plicator 1cm below the GE junction. The Plicator instrument remained in this position for 15 minutes, based on average procedure times established in previous studies. During the sham procedure, physician investigators were required to talk through the procedure steps to maintain the patient blind.
University of California at Irvine Medical Center
Orange, California, United States
Tri Valley Gastroenterology
San Ramon, California, United States
University of Colorado Health Science Center
Denver, Colorado, United States
Beth Israel Deaconess Medical Center
Boston, Massachusetts, United States
Creighton University Medical Center
Omaha, Nebraska, United States
Seacoast Gastroenterology
Exeter, New Hampshire, United States
Dartmouth Hitchcock Medical Center
Lebanon, New Hampshire, United States
Lenox Hill Hospital
New York, New York, United States
Columbia University Medical Center
New York, New York, United States
Nashville Medical Research and the Maria Nathanson Center of Excellence
Nashville, Tennessee, United States
...and 5 more locations
>/=50% improvement in GERD-Health Related Quality of Life Score
Time frame: 3-months post-procedure
GERD medication use
Time frame: 3-months post-procedure
Esophageal acid exposure
Time frame: 3-months post-procedure
Heartburn score as measure by visual analog scale
Time frame: 3-months post-procedure
Quality of Life as measure by SF-36
Time frame: 3-months post-procedure