The primary objective is to determine if patent foramen ovale (PFO) closure with the GORE® HELEX® Septal Occluder or GORE® CARDIOFORM Septal Occluder plus antiplatelet medical management is safe and effective and reduces the risk of recurrent stroke or imaging-confirmed transient ischemic attack (TIA) when compared to antiplatelet medical management alone in patients with a PFO and history of cryptogenic stroke or imaging-confirmed TIA. A co-primary objective is to demonstrate that medical management plus closure with the study device reduces the risk of new brain infarct compared to medical management alone.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
664
PFO closure with study septal occluder device
Investigator's choice of one of three regimen options specified in protocol
University of Pennsylvania Medical Center
Philadelphia, Pennsylvania, United States
Rigshospitalet
Copenhagen, Denmark
Number of Subjects With Freedom From Recurrent Ischemic Stroke (Primary Outcome #1)
A recurrent stroke event was defined as the first occurrence post-randomization of either a) neurological deficit presumed due to ischemia and persisting longer than 24 hours or until death, or b) transient neurological deficit presumed due to ischemia, persisting less than 24 hours with MRI evidence of a new relevant brain infarction.
Time frame: 24 months
Number of Subjects With New Brain Infarct or Recurrent Stroke (Primary Outcome #2)
Responders were subjects who showed one or more new infarctions on MRI since screening, or experienced a confirmed recurrent stroke, through 24 months (913 days). Nonresponders were subjects who did not show new infarction on MRI since screening and were confirmed free of recurrent stroke through at least 549 days. An infarction was defined as a new (since screening) T2 hyperintense MRI lesion with diameter ≥ 3 mm.
Time frame: 24 months
Number of Subjects With Study-related Serious Adverse Events
Adverse event seriousness and relationship to study treatments (device, procedure, or antiplatelet medical therapy) as reported by each investigative site
Time frame: 24 months
Number of Subjects With Effective Closure in Test (Device) Arm
Assessment of PFO closure in test (device) arm subjects by transesophageal echocardiography (TEE) at 24-month follow-up. Effective closure defined as occluded, small, or moderate shunt (0-25 bubbles).
Time frame: 24 months