This study is a prospective, non-randomized, multi-center clinical trial that will assess the safety and effectiveness of investigational models of the Promote™ RF CRT-D and Current™ RF ICD devices. These new ICD and CRT-D devices have new features that allow the device to periodically check how much energy is needed to pace a patient's heart and to automatically adjust the amount of energy used to pace the heart.
All patients enrolled in the study will have their device tested to determine if the device can automatically measure how much energy is needed to pace the patient's heart. If the test is successful, the patient's device will be programmed to turn that feature on. 3 months after enrollment, the patient will return to the clinic to have their device re-tested using an automatic test method to measure the amount of energy needed to make the patient's heart beat. A manual test method will also be used to measure the amount of energy needed to make the patient's heart beat. The intent of the study is to show the automatic method works just as well as the manual method.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
128
Patients receiving cardiac resynchronization therapy will have new autocapture features tested on right atrial, right ventricular and left ventricular leads.
Patients with an Implantable Cardioverter Defibrillator will have new autocapture features tested on the right atrial and right ventricular leads.
Arkanasas Heart Hospital
Little Rock, Arkansas, United States
Baptist Health Hospital
Little Rock, Arkansas, United States
Deborah Heart and Lung Center
Browns Mills, New Jersey, United States
St. Thomas Hospital
Nashville, Tennessee, United States
Percentage of Participants Free of System-related Complications at 3-months Post Implant
Time frame: 3 months post implant
Atrial AutoCapture (ACap) Confirm Effectiveness Endpoint - Absolute Difference Between the Automatic and Manual Capture Threshold Test
Atrial AutoCapture Confirm is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the atrial pacing voltage until it determines that the device is no longer capturing the atria. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the atrial pulse voltage and determines the atrial threshold by viewing the EKG.
Time frame: 3 months post implant
Ventricular Autocapture Effectiveness Endpoint - Absolute Difference Between the Automatic and Manual Capture Threshold Test
Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.
Time frame: 3 months post implant
Right Ventricular (RV) AutoCapture Effectiveness Endpoint - Difference Between the Automatic and Manual Capture Threshold Test
Right Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Southwest Texas Methodist Hospital
San Antonio, Texas, United States
Time frame: 3 months post implant
Left Ventricular (LV) AutoCapture Effectiveness Endpoint - Difference Between the Automatic and Manual Capture Threshold Test
Left Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.
Time frame: 3 months post implant