To demonstrate that ablation with the Therapy Cool Path Duo cardiac ablation system can eliminate ischemic VT and that its use does not result in an unacceptable risk of serious adverse events.
Ventricular Tachycardia (VT) is a common complication of ischemic heart disease that is often life threatening. Ventricular tachycardia is defined as three or more beats of ventricular origin in succession at a rate greater than 100 beats per minute. There are no normal QRS complexes associated with VT, and the rhythm is usually regular. Patients with VT experience unpleasant heart palpitations, dyspnea, and syncope, all of which decrease their overall quality of life. They also suffer the consequences of impaired hemodynamics, loss of cardiac efficiency, and are at an increased risk of sudden cardiac death (SCD), which accounts for approximately half of all deaths from patients with cardiovascular disease. Historically, antiarrhythmic medications served as the first line therapy for the treatment of VT, in spite of low efficacy and high recurrence rates. But in recent years, internal cardioverter defibrillators (ICDs) have become the treatment of choice. Although, ICDs are effective in terminating most arrhythmias, rarely does so without some episodes requiring poorly tolerated shock therapy. Ablative therapy offers an important option for frequent control with the potential for long-term VT elimination. Radiofrequency catheter ablation offers the ability to provide immediate control of recurrent VT. Since the early 2000's, strategies, tools, and techniques have been improving to identify key ablation sites and to deliver effective lesions. Radiofrequency ablation is successful in treating some but not all arrhythmias. With non irrigated technology it has been difficult to ablate ischemic VT. This may be due to an inadequate lesion size. Irrigated electrodes were developed to create larger lesions. Open irrigation at the catheter tip not only produces larger lesions, but also helps reduce the risk of coagulum (clots) and charring due to high temperatures.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
20
Cardiac Ablation Procedure for Ischemic VT using a flexible, insulated 7F all braided catheter that contains an internal lumen connected to 12 open conduits at the 4mm tip electrode for infusion of heparinized saline during radiofrequency ablation; records intracardiac electrograms and can be utilized for cardiac simulation during diagnostic electrophysiologic studies/evaluation. 1500T9-VT generator is a microprocessor-controlled device that produces a continuous unmodulated radiofrequency (RF) output of 485 kHz. The Generator will be used in Temperature Control mode only. The catheter delivers the RF power from the generator in a monopolar mode between its distal electrode (tip electrode) and a large indifferent electrode.
University Hospital - Univ. of Alabama at Birmingham (UAB)
Birmingham, Alabama, United States
Loyola University Medical Center
Maywood, Illinois, United States
Krannert Institute of Cardiology
Indianapolis, Indiana, United States
Kansas University Medical Center
Kansas City, Kansas, United States
Primary Safety is Defined as the Incidence of Intra-procedural, Acute or Sub-chronic, Serious Cardiac Adverse Events, up to 7 Days Post-procedure.
Time frame: 7 days
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Brigham and Women's Hospital
Boston, Massachusetts, United States