The primary objective of the study is to demonstrate the linear wear rates of the Trident® X3 polyethylene insert are superior to a polyethylene control group wear rate at 5 years postoperative. This measurement will be evaluated by comparing digitized images of serial radiographs obtained over a follow-up period of 5-years.
The study is a prospective, multi-center, historical-controlled clinical evaluation of the Trident® X3 Polyethylene insert. The device is commercially available in the United States where the study is being conducted. The study device, the Stryker Orthopaedics Trident® X3 polyethylene insert, is used for the replacement of the bearing surface of the acetabulum to relieve pain, instability, and the restriction of motion due to degenerative bone disease, including osteoarthritis, rheumatoid arthritis, trauma, or failure of other devices. The control device, the Stryker Orthopaedics N2VAC polyethylene insert, is also used for the replacement of the bearing surface of the acetabulum to relieve pain, instability, and the restriction of motion due to degenerative bone disease, including osteoarthritis, rheumatoid arthritis, trauma, or failure of other devices.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
250
Trident® X3 Polyethylene Insert
Emory University
Atlanta, Georgia, United States
Bonutti Research Inc.
Effingham, Illinois, United States
Indiana University
Indianapolis, Indiana, United States
New England Baptist Hospital
Mean Linear Wear Rate at 5 Years
Linear wear rates are defined as the annual rate of removal of the polyethylene from the polyethylene insert determined by comparing digitized images of serial radiographs obtained over the follow-up period of 5 years
Time frame: 5 years
Rate of Incidence of Revision of Component for Any Reason
Revision of any component is defined as surgical removal and replacement of the femoral component, acetabular shell, acetabular insert and/or femoral head.
Time frame: 5 year
Linear Wear Rate of the Trident X3 Polyethylene Insert
Linear wear rates are defined as the annual rate of removal of the polyethylene from the polyethylene insert determined by comparing digitized images of serial radiographs obtained over the follow-up period. NOTE: 2 year linear and volumetric wear was not calculated for the following reason: To determine polyethylene wear, the total femoral head penetration is first calculated from the radiographs.The femoral head penetration has two components namely wear and creep (or bedding-in). It is not possible to separate the total penetration in to two components. The creep of the polyethylene starts from the date of surgery and continues up to 12-24 months. Therefore, the head penetration value at 2-years is dominated by Creep rather than wear.
Time frame: 2, 3 and 4 year films collected; 3 and 4 year wear assessed
Volumetric Wear Rate of the Trident X3 Polyethylene Insert
Volumetric wear rate is calculated using a formula based on the cylindrical wear pattern perpendicular to the face of the cup and the mean linear wear rate. NOTE: 2 year linear and volumetric wear was not calculated for the following reason: To determine polyethylene wear, the total femoral head penetration is first calculated from the radiographs.The femoral head penetration has two components namely wear and creep (or bedding-in). It is not possible to separate the total penetration in to two components. The creep of the polyethylene starts from the date of surgery and continues up to 12-24 months. Therefore, the head penetration value at 2-years is dominated by Creep rather than wear.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Boston, Massachusetts, United States
J. Wesley Mesko, MD
Lansing, Michigan, United States
Mayo Clinic
Rochester, Minnesota, United States
Good Samaritan Hospital
Kearney, Nebraska, United States
Greater Pittsburgh Orthopaedics Assoc
Pittsburgh, Pennsylvania, United States
University of Vermont College of Medicine
Burlington, Vermont, United States
Time frame: 2, 3, 4 and 5 year films collected; 3, 4 and 5 year wear assessed
Radiographic Stability
Radiographic stability is defined as having all of the following: no radiographic indication of progressive radiolucent lines greater than or equal to 2 mm around the entire acetabular cup, no radiographic indication of acetabular cup migration of greater than or equal to 3 mm, no radiographic indication of progressive radiolucent lines greater than or equal to 2 mm around the entire femoral component, and no radiographic indication of progressive subsidence of the femoral component of greater than or equal to 5 mm. Radiographs are evaluated at 1,2,3,4 and 5 years.
Time frame: 1,2,3,4 and 5 years
Mean Harris Hip Score (HHS) to Assess Change
The change in HHS is reported by comparing the mean preoperative, 1, 3 and 5 year postoperative scores that assess pain, function, joint deformity and range of motion. Scores can range from 0 to 100 with 0 being the worst and 100 being the best score. A score of 80-100 is considered good-excellent and a score less than or equal to 79 is considered fair-poor. 90 - 100 = excellent 80 - 89 = good 70 - 79 = fair 0 - 69 = poor
Time frame: pre-operative, 1,3 and 5 years
Mean Harris Hip Score (HHS) Pain Score to Assess Change
The change in HHS Pain is reported by comparing the mean preoperative, 1, 3 and 5 year postoperative pain scores. Scores can range from 0 to 44, with 0 indicating totally disabling pain and 44 indicating no pain or pain that is ignored. * None or ignores it = 44 points * Slight, occasional, no compromise in activities = 40 points * Mild pain, no effect on average activities, rarely moderate pain with unusual activity;may take aspirin = 30 points * Moderate pain, tolerable, but makes concessions to pain. Some limitation of ordinary activity or work. May require occasional pain medication stronger than aspirin = 20 points * Marked pain, serious limitation of activites = 10 points * Totally disabled, crippled, pain in bed, bedridden = 0 points
Time frame: pre-operative, 1,3, and 5 years
Mean Harris Hip Score (HHS) Range of Motion (ROM) Score to Assess Change
The change in HHS ROM is reported by comparing the mean preoperative, 1, 3 and 5 year postoperative scores. Scores can range from 0 (worst) to 5 (best). The degrees of motion are measured for hip flexion, abduction, adduction, external rotation and internal rotation. The measured values are added to determine a combined value that is associated with a score from 0 to 5. * 211-300 degrees = 5 points * 161 to 210 degrees = 4 points * 101 to 160 degrees = 3 points * 61 to 100 degrees = 2 points * 31 to 60 degrees = 1 points * 0 to 30 degrees = 0 points
Time frame: pre-operative, 1,3 and 5 years
Mean SF-12 Health Survey Score to Assess Change
Change in the SF-12 score is reported by comparing the mean preoperative, 1,3 and 5 year postoperative scores.The SF-12 Health Survey is a 12-item patient completed questionnaire to measure general health and well-being. It includes a physical and mental status component score; each ranging from 0-100. Low values represent a poor health state and high values represent a good health state.
Time frame: pre-operative, 1,3 and 5 years
Mean Lower Extremity Activity Scale (LEAS) Score to Assess Change
Change in the LEAS is reported by comparing the mean preoperative, 1,3 and 5 year scores. The LEAS is completed by the participant to assess activity level. Activity levels are ordered in terms of intensity from 1 to 18, with 18 indicating the highest activity level.
Time frame: pre-operative, 1,3 and 5 years
Implant Survivorship
Implant survivorship is determined using the Kaplan-Meier method.
Time frame: 10 years