The purpose of this study was to compare the Continuous Non-invasive Arterial Pressure (CNAP) monitor with the gold standard of invasive arterial pressure monitoring during: 1. induction and maintenance of general anaesthesia 2. intensive care unit treatment of postoperative patients with an ASA 3 or 4 status
The CNAP monitor continuously measures blood pressure using a finger cuff. It showed a good agreement to invasive blood pressure measurements during anaesthesia induction and maintenance during surgery. In critical ill patients cardiac arrhythmia is probably a confounding factor affecting accuracy and interchangeability of CNAP. As systolic arterial CNAP pressure in comparison to mean pressure shows no statistical interchangeability with invasive measurements, mean pressure should be considered when making therapy decisions. A problem is the missing standard criterion for comparison of continuously devices with invasive pressure.
Study Type
OBSERVATIONAL
Enrollment
195
Klinik für Anästhesiologie und Operative Intensivmedizin, Universitätsklinik Schleswig-Holstein, Campus Kiel
Kiel, Germany
Bias and percentage error of CNAP compared with invasive arterial pressure measurement
Time frame: 1.) Induction of general anaesthesia. 2.) Maintanance of general anaesthesia. 3.) Intensive Care Treatement after surgery