The proposed investigation is a prospective, multicenter, non-randomized study to evaluate the immediate benefits (at discharge and at 30 days) in terms of performance and safety of the implantation of the Medtronic CoreValve PAV in a failing aortic bioprosthesis. In addition, the study aims to evaluate the performance and the safety of the implantation of the Medtronic CoreValve PAV in a failing aortic bioprosthesis at subsequent annual follow-ups out to 48 months post procedure. These objectives will be achieved through the following endpoints: * Primary safety endpoint - Composite of Major Adverse Events * Primary performance endpoint - Technical and procedural success at discharge Patient Population: Eligible subjects will be at least 75 years old, presenting with a failing aortic bioprosthesis (stenotic, incompetent or mixed) including homograft or stented or stentless heterograft, considered poor surgical candidates and necessitating repeat aortic valve replacement. Up to 20 patients will be included in up to four hospitals.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
18
Transcatheter Aortic Valve
Herzzentrum Leipzig GmbH
Leipzig, Germany
Deutsches Herzzentrum München
München, Germany
Erasmus MC
Rotterdam, Netherlands
Percentage of MAE Free Subjects
Percentage of subjects without a composite major adverse event (MAE) at 30 days post procedure. the composite MAE includes: 1. MACCE (Major Adverse Cardiac or Cerebral Event): cardiac death, documented Mobitz type II second degree and third degree atrioventricular block, neurological events and surgical aortic valve replacement. 2. MAE: non cardiac death including sudden unexpected or unexplained death, cardiac tamponade, nonstructural valve dysfunction resulting in stenosis or regurgitation at the study valve not intrinsic to the valve itself, structural valve deterioration cardiogenic shock, operated valve endocarditis, valve thrombosis, aortic dissection/perforation, major bleeding event, peripheral embolic event and emergent revascularization procedure.
Time frame: 30 days post-procedure
Technical Success.
The number of subjects in whom technical success was achieved where technical success was defined based on the device functionality as judged by the investigator adressing the ability of the system (1) to access the failing bioprosthetic valve via a peripheral vessel with the delivery catheter and (2) to deploy the valve accurately across the failing bioprosthetic valve and (3) to remove intact delivery system.
Time frame: 30 days post procedure