Background: In patients with obstructive sleep apnea, automatic continuous positive airway pressure machines (autoCPAP) are said to be capable of identifying various breathing abnormalities during sleep and to correct them by increasing progressively the positive pressure applied to the airway. Once breathing becomes normal, pressure slowly declines. AutoCPAP devices have never been tested in Phase I studies. The investigators hypothesised that normal breathing would not be recognised as such, and that pressure would increase even in a normal subject.
Material and Methods: We will submit one normal (confirmed with two polysomnographies, PSG) subject to a double blind study. Pairs of PSG will be performed on successive days once a week, one night with a 4 cm water fixed pressure CPAP device, previously shown not to disturb the subject's sleep and breathing, the other night with one of five well known autoCPAP devices programmed to work between 4 and 15 cm H2O pressure. The same mask and headgear will be used throughout the study. The subject and the physicians reading the tracings will ignore which device is being used on which night.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
QUADRUPLE
Enrollment
1
Cliniques universitaires Saint-Luc
Brussels, Belgium
number of episodes of increases in applied positive pressure, duration of each episode, and maximal pressure attained
Time frame: one night
reasons for pressure increase and consequences of the increase on sleep continuity
Time frame: one night