The purpose of this study is to demonstrate diagnostic agreement and determine the accuracy of the continuous positive airway pressure (CPAP) device compared to simultaneous, attended clinical polysomnography (PSG) in identifying breathing events in participants previously diagnosed with complex sleep apnea (CompSAS), complex sleep apnea with Cheyne-Stokes respiration (CSR), or obstructive sleep apnea (OSA).
Study Objectives: To compare a positive airway pressure (PAP) device's detection of respiratory events and airway status during device detected apneas with events scored on simultaneous polysomnography (PSG). Design: Prospective PSGs of patients with sleep apnea using a new-generation PAP device. Settings: Four clinical and academic sleep centers. Patients: Forty-five patients with obstructive sleep apnea (OSA) and complex sleep apnea (Comp SA) performed a PSG on PAP levels adjusted to induce respiratory events. Interventions: None.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
45
Positive airway pressure (PAP) will be manipulated throughout the night to induce breathing events. PAP will be set to the participant's prescribed pressure, increased until breathing events are induced , and then returned to the prescribed pressure. This cyclic pattern will continue throughout the night.
National Jewish Health
Denver, Colorado, United States
University of Florida - Shands Sleep Disorders Center
Gainesville, Florida, United States
Sleep Health
Portage, Michigan, United States
Number of Breathing Events Identified by the Continuous Positive Airway Pressure (CPAP) Device Compared to a Simultaneous Polysomnography
The following breathing events: RERAs, central apneas, periodic breathing, obstructive apneas, and hypopneas in patients previously diagnosed with CompSAS or OSA detected by simultaneous Polysomnography and REMstar Auto with A-Flex were compared.
Time frame: During a single night of polysomnography lasting an average of 8 hours
Device Detected Apneas as Detected by Philips Respironics (PR) System One
All device-detected apneas were tabulated. Then they were broken down into obstructed airway apneas and clear airway apneas. The results are recorded below as apneas with obstructed airway and apneas with clear airway.
Time frame: During a single night of polysomnography lasting an average of 8 hours
Device-Detected Obstructed Airway Apnea Agreement
Device-Detected Apneas were broken down and then reviewed with the manually scored apneas. Of the Obstructed Airway apneas they were broken down into a few different categories: Manually Scored Obstructive Apneas, Manually Scored Central Apneas, Manually Scored Hypopneas and Manually Scored RERAs.
Time frame: During a single night of polysomnography lasting an average of 8 hours
Device-Detected Clear Airway Apnea Agreement
Device Detected Clear Airways events were counted by the device. These events were then compared to manual PSG scoring. The manual PSG scoring determined that these were apnea events. However, the manual scoring allows for more channels to be reviewed and manual scoring was able to classify them into 4 different categories: Manually Scored Obstructive Apneas, Manually Scored Central Apneas, Manually scored hypopneas and manually scored RERAs.
Time frame: During a single night of polysomnography lasting an average of 8 hours