It's the aim of this study to analyse efficacy and reliability of VectSelect Feature for avoidance of phrenic nerve stimulation in cardiac resynchronisation therapy (CRT) patients.
Study Type
OBSERVATIONAL
Enrollment
132
Medizinische Einrichtungen der RWTH Aachen
Aachen, Germany
Hufeland Klinikum GmbH
Bad Langensalza, Germany
Cardioangiologisches Centrum Bethanien
Frankfurt am Main, Germany
Amount of patients with at least one acceptable LV pacing configuration.
Amount of patients with at least one acceptable, programmable LV pacing configuration. Definition of acceptable LV pacing configuration: LV pacing threshold \<=2.5V@0.5ms AND Phrenic nerve threshold is at least twice as high as LV pacing threshold AND LV pacing impedance \>=200 Ohms. Only patients with implanted CRT-D and bipolar leftventricular lead are taken into account.
Time frame: First Follow Up after hospital discharge (approx. 1-3 months)