The purpose of this study is to evaluate the performance and acceptability of an investigational blood glucose monitoring system, which includes meter and sensor strip.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
94
Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter and an investigational sensor. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff.
AMCR Institute
Escondido, California, United States
Numbers of Fingerstick Blood Glucose (BG) Results Within +/- 5 to 20mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method
Subjects with diabetes tested subject fingerstick blood using an investigational blood glucose monitoring system (BGMS), which included an investigational meter and sensor. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results \<75mg/dL) or +/- 5 to 20% (for reference BG results \>=75mg/dL) of the reference method results.
Time frame: 2 hours
Numbers of Palm Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method
Subjects with diabetes used an investigational blood glucose monitoring system (BGMS) with subject capillary blood obtained from the palm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results \<75mg/dL) or +/- 5% to 20% (for reference BG results \>=75mg/dL) of the reference method results.
Time frame: 2 hours
Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension)
Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects (1 to 4) on their success at performing the tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required.
Time frame: 2 hours
Numbers of Forearm Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Subjects with diabetes used an investigational blood glucose monitoring system (BGMS) with subject capillary blood obtained from the forearm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results \<75mg/dL) or +/- 5% to 20% (for reference BG results \>=75mg/dL) of the reference method results.
Time frame: 2 hours