Impact of intraventricular electrical activation in resynchronization therapy. We seek to evaluate the impact of Cardiac Resynchronization Therapy (CRT) on electrical activation of the Left Ventricle (LV). The first goal of the study is to evaluate if CRT is able to decrease the heterogeneity of LV activation in heart failure patients. A second goal is to evaluate the electrical determinant of clinical response to CRT using invasive and non-invasive mapping technology.
Comprehensive study, prospective, monocentric and cohort study. Biventricular pacing is an effective therapy to improve end stage heart failure patients, however up to 30% of patients are unresponsive to this therapy. Among others, optimization of Left Ventricle (LV) pacing site location is critical to optimize response to Cardiac Resynchronization Therapy (CRT). Whether a better comprehension of baseline LV electrical activation can help to determine optimal LV lead location remains to be investigated. After informed consent patients are included in the sudy. Day 1: CRT device is implanted under general anesthesia in a conventional fashion. The coronary sinus leads will be implanted guided by the acute hemodynamic response (+(dP/dt)max) using a RADI guide wire. After device implantation, a steerable mapping catheter will be inserted in the LV cavity to build a tridimensional activation map of the LV. This map will be obtained with and without CRT. From these maps different parameters describing LV activation will be measured and compared between patients with and without CRT. Quantitative parameters: Global LV activation time (LVATG) (primary outcome); septal activation time (LVTAS); Global intra-LV dispersion time and Qualitative parameters: presence of intraLV line of block, area of late activation. Day 5±2: A non-invasive map of the LV will be recorded with the CardioInsight ® system with and without CRT. The same parameters will be measured and compared with those obtained with invasive mapping. A 3 and 6 months follow-up visit including clinical, biological evaluation (NYHA status, 6-min walking test, BNP) of the patients will be performed with an additional transthoracic echocardiography at 6 months.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Day 1: CRT device is implanted under general anesthesia in a conventional fashion. The coronary sinus leads will be implanted guided by the acute hemodynamic response (+(dP/dt)max) using a RADI guide wire. After device implantation, a steerable mapping catheter will be inserted in the Left Ventricle (LV) cavity to build a tridimensional activation map of the LV. This map will be obtained with and without CRT. From these maps different parameters describing LV activation will be measured and compared between patients with and without CRT. Day 5±2: A non-invasive map of the LV will be recorded with the CardioInsight ® system with and without CRT. The same parameters will be measured and compared with those obtained with invasive mapping.
Hôpital cardiologique du Haut-Lévêque, Pr. Haïssaguerre's department, 3rd Floor
Pessac, France
Evolution of the Global Left Ventricle activation time (LVATG) with and without Cardiac Resynchronization Therapy (CRT).
Time frame: Day 1 : End of CRT device implantation
Number, area and location of line of conduction block in the Left Ventricle
Time frame: Day 1 : End of CRT device implantation
Area of the latest activated region of the Left Ventricle
Time frame: Day 1 : End of CRT device implantation
Number, area and location of LV region activated after the end of surface QRS
Time frame: Day 1 : End of CRT device implantation
Septal activation time (LVTAS)
Time frame: Day 1 : End of CRT device implantation
Global intra-Left Ventricular dispersion time
Time frame: Day 1 : End of CRT device implantation
Determination of the electrical parameters associated with good clinical response to CRT at 6 months.
Time frame: 6 months after CRT device implantation
Non-invasive mapping: Same criteria evaluated non-invasively using CardioInsight ® with and without Cardiac Resynchronization Therapy activated.
Time frame: Between 3 to 7 days after CRT device implantation
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Masking
NONE
Enrollment
18