Purpose: : This study was performed to compare the accuracy of non-cycloplegic photorefraction with cycloplegic refraction in detection of refractive amblyopia risk factors and determine some corresponding beneficial cut points. Method: In this diagnostic accuracy study, right eyes of 185 children (1 to 14 years) underwent first non-cycloplegic photorefraction (with PlusOptix SO4 photoscreener) and then cycloplegic refraction from October 2009 to August 2010. Based on the cycloplegic refraction results, hyperopia (≥+3.5D), myopia (≥-3D), astigmatism (≥1.5D), and anisometropia (≥1.5D) were set as criteria based on AAPOS guidelines. The difference in detection of refractive amblyopia risk factors between the two methods was the main outcome measure.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
DOUBLE
Enrollment
185
photorefraction is a name of device.
Labbafinejad medical center
Tehran, Iran
refraction
cycloplegic auto refraction \& Non-cycloplegic photorefraction
Time frame: 10 months
hyperopia (≥+3.5D), myopia (≥-3D), astigmatism (≥1.5D), and anisometropia (≥1.5D) were set as criteria based on AAPOS guidelines.
cycloplegic auto refraction \& Non-cycloplegic photorefraction
Time frame: 10 months