This study evaluates the ability of a prototype intraoperative handheld gamma camera (pIHGC) to image (detect) sentinel lymph nodes (SLNs) in melanoma and breast cancer during surgical excision, as compared to standard of care intraoperative gamma probes (GP). The unit of study in this trial was SNLs rather individual participants. Each device was assessed for relative node detection sensitivity (S) of those SLNs.
Lymphoscintigraphy is an accepted and commonly-performed procedure used for staging of certain cancers, especially melanoma and breast cancer. It involves injecting a small amount of radioactivity under the skin in order to identify lymph nodes which should be biopsied (ie, the "sentinel lymph node, SLN") to determine if cancer has spread. The study objective was to evaluate the potential benefit of a new, camera-based technology (prototype device) which allows actual images to be obtained intra-operatively in the identification of sentinel nodes. Each device was assessed for relative node detection sensitivity (S), defined as the proportion of sentinel lymph nodes (SLNs) that were identified with each instrument, with the proportion determined as the number of true positives (TP) divided by the total evaluated (N).
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
50
The prototype intraoperative handheld gamma camera (pIHGC) consists of a parallel-hole lead collimator coupled to a pixilated sodium iodide-thallium \[NaI(Tl)\] scintillation crystal array, itself coupled to a flat panel, multi-anode Hamamatsu H8500 position-sensitive photomultiplier tube. The collimator is 5x5 cm2 large in area and 1.5 cm thick, with 1.3 mm hexagonal holes and 0.2 mm septa. The 1.7 mm pitch crystal array is composed of 29x29 individual crystals, each 1.5x1.5x6 mm3 in size.
Lymphoscintigraphy with standard of care intraoperative gamma probes (GP)
Stanford University School of Medicine
Stanford, California, United States
Relative Node Detection Sensitivity
Relative node detection sensitivity (S) was defined as the proportion of sentinel lymph nodes (SLNs) that were identified with each instrument, with the proportion determined as the number of true positives (TP) divided by the total evaluated (N). The outcome result is expressed as the percentage for each device with 95% confidence intervals.
Time frame: 1 day
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Lymphoscintigraphy involves injection of 0.4 to 1.0 mCi of radioactive Tc99M sulfur colloid around at the tumor site.