The study evaluated the safety and efficacy of the vBloc Therapy delivered by the Maestro Rechargeable System compared to a sham control for the treatment obesity. The Maestro Rechargeable System delivers intermittent, electrical blocking signals to the anterior and posterior trunks of the intra-abdominal vagus nerve (termed vBloc Therapy) and is designed to reduce sensations of hunger and produce satiety leading to weight loss. After approval from the institutional review board, subjects provided written consent and were randomized in a 2 to 1 allocation to vBloc group (with laparoscopic placement of the leads and neuroregulator) or the sham group (placement of a custom sham neuroregulator only) on double-blinded basis. Both groups received similar diet and exercise counseling. Weight, adverse events, clinical laboratories, ECGs, eating questionnaires and quality of life data were evaluated throughout the 12 month study duration.
The Maestro® Rechargeable System is a neuromodulation system. All subjects will be randomized in a 2:1 allocation to vBloc or Sham. All subjects will be followed through the 12-month follow-up visit. Subjects in the Sham group who choose to have an active device will undergo a second surgery to receive the active device after study unblinding.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
DOUBLE
Enrollment
239
Active device will deliver vBloc Therapy
Functional non-active control device will deliver no vBloc Therapy
HonorHealth (formerly Scottsdale Healthcare Bariatric Center)
Scottsdale, Arizona, United States
Scripps Clinic Nutrition & Metabolic Research Center
La Jolla, California, United States
Stanford University
Stanford, California, United States
Tufts Medical Center
Boston, Massachusetts, United States
University of Minnesota
Minneapolis, Minnesota, United States
Mayo Clinic Rochester
Rochester, Minnesota, United States
Oregon Health & Science University
Portland, Oregon, United States
Virginia Commonwealth University
Richmond, Virginia, United States
Institute of Weight Control
Baulkham Hills, New South Wales, Australia
Adelaide Bariatric Centre
Bedford Park, SA, Australia
Percentage of Subjects Experiencing Implant/Revision Procedure, Device or Therapy Related Serious Adverse Events (SAEs).
To demonstrate that the implant/revision procedure, device and therapy related serious adverse event rate in the vBloc group at 12 months post-implant is significantly lower than 15%.
Time frame: 12 months
Percentage of Excess Weight Loss (EWL) by Body Mass Index (BMI) Method.
Observe at least a 10% greater excess body weight loss (EWL) from randomization with the Maestro System after 12 months of vBloc Therapy compared to Sham by body mass index (BMI) method. (Body mass index is calculated by dividing body weight (kg) by body height (m) squared (BMI=kg/m2)).
Time frame: 12 months
Percentage Responder Rate in the Treatment Arm.
The second co-primary effectiveness endpoint was based on responder rates with the following two requirements: (i) at least 55% of vBloc subjects would achieve a %EWL of at least 20%; and (ii) at least 45% of vBloc subjects would achieve a %EWL of at least 25%.
Time frame: 12 months
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.