Specific objectives include confirmation: i) that the ROM PLUS consistently and accurately diagnoses rupture of membranes (ROM), ii) that the technique can be understood and applied in clinical practice, and iii) that the ROM PLUS is easier to use than the conventional method.
This study is designed to assess the sensitivity and specificity of ROM PLUS- a rapid, point of care, qualitative immunochromatographic test for the detection (in vitro) of amniotic fluid in cervico-vaginal secretions of women with suspected rupture of membranes (ROM) during pregnancy. The ROM PLUS detects a specific Combo protein present in amniotic fluid of pregnant women in all trimesters of pregnancy. This specific protein combo is unique and found only in amniotic fluid, therefore can be used as a specific marker for the diagnosis of ROM. This biomarker is an isoform of a similar biomarkers called Amni-Sure and ActimProm and was recently approved by the FDA for clinical use. Like these protein markers ( AmniSure or ActimProm) which have been used to diagnose PROM, this particular protein isoform tested for ROM PLUS is in low concentration in other body fluids such as maternal blood, cord blood, urine and semen.
Study Type
OBSERVATIONAL
Enrollment
288
Sterile vaginal speculum exam to include Nitrazine, ferning and pooling observations
Vaginal swab exam for ROM Plus Test
Post delivery blinded chart review by expereienced obstetrician
University of Utah Medical Center
Salt Lake City, Utah, United States
Pregnant Women Positive and Negative for Membrane Rupture Measured Via Clinical Assessment, Chart Review and ROM Plus
Patients underwent two assessments to determine positive or negative membrane rupture status: (1) Standard clinical assessment using fluid leaking from the cervical os, or two of the following; pooling, positive nitrazine test, or ferning and (2) A new combination immunoassay ROM Plus containing a combination of monoclonal and polyclonal antibodies to Placental Protein 12 (PP12) and Alpha-fetoprotein (AFP). Then, membrane rupture status was determined by chart review for reference based on a post delivery patient chart review by an experienced physician blinded to ROM Plus results.
Time frame: 1 week
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.