The purpose of the IVORY Study is to gather data to support global submissions/approvals for some models of the Ingenio device family.
IVORY is a prospective, multi-center, randomized within-patient, single-blinded study to gather data to support Right Atrial Auto Threshold and RYTHMIQ
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
139
Wilhelminenspital
Vienna, Austria
Landesklinikum Wiener Neustadt
Wiener Neustadt, Austria
Accuracy of Commanded Right Atrial Automatic Threshold (RAAT)
Accuracy was evaluated by the proportion of tests resulting in an RAAT commanded threshold with \|Manual unipolar threshold - commanded threshold\| ≤ 0.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for purposes of endpoint analysis. A paired dataset consisted of the manual unipolar threshold and the commanded threshold obtained at a visit.
Time frame: 3 months post implant
System-related Complication-free Rate
Arms/Groups are only applicable for the outcome measure 'Secondary Outcome: RYTHMIQ - Median Relative Reduction of RV Pacing Percent (RVPP) for RYTHMIQ Programmed to ON vs. RYTHMIQ Programmed to OFF'. This outcome measure was analyzed only for the combined population and both pacemakers and CRT-P devices were pooled for the analysis. The analysis assesses the system-related complication (SRC) -free rate (excluding LV related events for CRT-P subjects) for subjects implanted or attempted to be implanted with a study device through the first 90 days following the implant procedure.
Time frame: 90 days post-implant
Accuracy of Ambulatory RAAT
Arms/Groups are only applicable for the outcome measure 'Secondary Outcome: RYTHMIQ - Median Relative Reduction of RV Pacing Percent (RVPP) for RYTHMIQ Programmed to ON vs. RYTHMIQ Programmed to OFF'. This outcome measure was analyzed only for the combined population (pacemaker and CRT-P). Accuracy was evaluated by the proportion of tests resulting in an RAAT ambulatory threshold with \|ambulatory threshold - commanded threshold\| ≤ 1.0V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for purposes of endpoint analysis. A paired dataset consisted consisted of the ambulatory threshold obtained within 7 days prior to the visit and the commanded threshold obtained at the visit.
Time frame: 3 months post-implant
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Clinique Universitaires Saint Luc
Brussels, Belgium
Rigshospitalet Copenhagen
Copenhagen, Denmark
Gentofte University Hospital
Hellerup, Denmark
NCN Nouvelles Cliniques Nantaises
Nantes, France
Clinique Saint-Hilaire Rouen
Rouen, France
Krankenhaus Neu Bethlehem
Göttingen, Germany
Heinrich Braun Krankenhaus
Zwickau, Germany
Prince of Wales Hospital
Shatin, New Territories, Hong Kong
...and 11 more locations
Appropriate RAAT Test Outcome
Arms/Groups are only applicable for the outcome measure 'Secondary Outcome: RYTHMIQ - Median Relative Reduction of RV Pacing Percent (RVPP) for RYTHMIQ Programmed to ON vs. RYTHMIQ Programmed to OFF'. This outcome measure was analyzed only for the combined population (pacemaker and CRT-P). Accuracy was evaluated by the proportion of tests with an appropriate commanded RAAT test outcome. An appropriate RAAT outcome consists of either a device-determined threshold or a code indicating that a threshold could not be determined and representing a condition that is beyond the control of the RAAT feature and could occur as well in manual threshold test, such as atrial fibrillation.
Time frame: 3 months post-implant
RYTHMIQ - Median Relative Reduction of RV Pacing Percent (RVPP) for RYTHMIQ On at Pre-Discharge, Off at 1-Month vs. RYTHMIQ Off at Pre-discharge, On at 1-Month
Chronic success is assessed by a median relative reduction of RV pacing percent (RVPP) for RYTHMIQ On at Pre-Discharge, Off at 1-Month vs. RYTHMIQ On at Pre-Discharge, Off at 1-Month. All patients received both RYTHMIQ programmed to ON and programmed to OFF in two consecutive periods, where the patient's programming allocation was randomized for the first period. Then a crossover to the alternate setting occurred for the second period once the end of the first period was reached.
Time frame: 3 months post implant
Percentage of RAAT Threshold Tests With Sufficient RAAT Pace Output Margin
This outcome measure was analyzed only for the combined population (Pacemaker and CRT-P). The sufficient pacing output voltage that is able to capture the right atrium was evaluated by the proportion of Right Atrial Automatic Threshold (RAAT) tests with \[RAAToutput ≥ (max(BP,UP) + δ )\] where BP refers to the manual bipolar threshold, UP refers to the manual unipolar threshold, and δ refers to the safety margin of 0.5V. Subjects were allowed to contribute multiple paired datasets for this endpoint analysis, one set each from the 1-month and 3-month visits. Paired datasets from the 1-month and 3-month visits were pooled for purposes of endpoint analysis. A paired dataset consisted of the commanded threshold and the maximum manual threshold (i.e. the maximum of the manual unipolar and manual bipolar thresholds) obtained at a visit.
Time frame: 3 months post implant