The objective of this Phase II study is to assess the diagnostic accuracy of sentinel lymph node (SLN) identification in patients with breast cancer using near-infrared (NIR) fluorescence optical imaging.
Fifty participants consented from Brigham and Women's Hospital (currently enrolling) and Beth Israel Deaconess Medical Center (currently non-enrolling), who will be undergoing standard of care for breast cancer SLN mapping with Tc-99m will be enrolled in this study. Prior to surgery, patients will undergo lymphoscintigraphy as is standard of care. At the time of SLN mapping, the patient will be injected with indocyanine green dye (ICG). Concurrently, a custom-designed imaging platform that utilizes NIR fluorescence optics will be employed to identify the dye and its path through the lymphatic track, and its eventual highlighting of the SLN. Both color images and NIR images will be merged to give the operating surgeon a clear picture of the ICG track superimposed over anatomical landmarks. This technique will identify the SLN and provide an accurate video image of its location. The surgeon will resect all SLNs identified by Tc-99m or NIR fluorescence, and then confirm the radioactivity and NIR fluorescence signal in each after resection.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DEVICE_FEASIBILITY
Masking
NONE
Enrollment
50
A custom-designed imaging platform that utilizes NIR fluorescence optics will be employed to identify the injected dye and its path through the lymphatic track, and its eventual highlighting of the SLN. Both color images and NIR images will be merged to give the operating surgeon a clear picture of the ICG track superimposed over anatomical landmarks. This technique will identify the SLN and provide an accurate video image of its location.
Dana Farber Cancer Institute
Boston, Massachusetts, United States
Sensitivity of Real-Time Intraoperative NIR Mapping
To determine the sensitivity of real-time intraoperative NIR lymphatic mapping in the identification of the sentinel lymph node in human breast cancer as compared to the standard of care, Technicium 99 (Tc-99).
Time frame: One Day (day 1)
Number of Participants with Adverse Events as a Measure of Safety and Tolerability
To evaluate the incidence of significant adverse reactions due to both NIR lymphatic mapping and indocyanine green.
Time frame: One Day (day 1)