The purpose of this study is to determine whether changes in a patient's position affect LV capture thresholds in a clinically significant way.
Traditionally clinicians program the safety margins for the right atrial, right ventricular, and left ventricular leads two-times the capture thresholds. This safety margin is necessary for the right atrial and right ventricular outputs to prevent asystole during fluctuation of capture thresholds. This two-times safety margin may not be necessary for the left ventricular lead output and may be causing shorter device battery longevity. Newly implemented LV capture threshold management algorithms may help maintain a proper balance of safety margin and battery longevity. However, recent research conducted on automatic left ventricular capture measurement (LVCM) shows as much as 3.5 V daily variability in capture thresholds. In this same study, 18% of patients had \>1.5 V LV threshold variability. A possible explanation for the variability of LV thresholds is lead stability. Small movements of the lead electrodes associated with movement of patient posture may account for daily fluctuation of capture thresholds. As heart failure patient care and therapies improve, CRT patients may survive longer than the projected longevities of their device. In order to minimize patient risks associated with device change-outs, care must be taken to maximize battery longevity while maintaining CRT pacing. Examining patient posture changes as a possible cause of threshold variability may guide clinicians to program more appropriate LV outputs with patient safety and device longevity in mind.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Masking
NONE
Enrollment
11
1:1 randomization between Medtronic active fixation 4195 LV lead and another FDA approved passive fixation Medtronic LV lead. Fluoroscopic images will be taken at implant in a supine position and again at 3 month follow-up in different postural positions with capture thresholds of the LV lead taken at each postural position.
The Heart Hospital in Plano
Plano, Texas, United States
Evidence of Change in Threshold of >1 Volt With Posture Changes
Time frame: 6 months
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.