The purpose of this study is to determine if untrained subjects who have diabetes can operate the Blood Glucose Monitoring System (BGMS) and obtain acceptable glucose results.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
102
Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick blood and AST of the palm and forearm using the Contour TS Blood Glucose Monitoring System. All BG results are compared to a reference laboratory glucose method.
Bayer HealthCare LLC, Diabetes Care
Mishawaka, Indiana, United States
Number of Self-Test Fingerstick Blood Glucose Results Within +/- 15mg/dL (<100mg/dL) or Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
Untrained subjects with diabetes self-test fingerstick blood using the Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Self-Test Alternative Site (Palm) Blood Glucose Results Within +/- 15mg/dL (<100mg/dL) or Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using the Blood Glucose Monitoring System (BGMS). BGMS AST palm results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST palm BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Self-Test Alternative Site (Forearm) Blood Glucose Results Within +/- 15mg/dL (<100mg/dL) or Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method
Untrained subjects with diabetes self-test Alternative Site (AST) Forearm blood using the Blood Glucose Monitoring System (BGMS). BGMS AST forearm results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST forearm BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Subject Responses That 'Agree' or 'Strongly Agree' With Questionnaire Statements
Subjects will complete short questionnaires to provide feedback on the labeling materials and system ease of use. Subjects may respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' 'Strongly Disagree'.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: 1 hour