The objective of the study is to evaluate the performance and intended use of a new cardiac biomarker test, Troponin I, in an intended use population. Blood specimens will be tested using the new investigational test that detects the level of Troponin I. Results will be compared to the diagnosis of whether or not an acute myocardial infarction (MI) occurred.
The purpose of the cardiac biomarker Troponin I is to aid in the diagnosis of myocardial infarction.The assay is also intended to assist in the prognosis relative risk to all cause mortality (ACM) and major adverse cardiac events (MACE) consisting of myocardial infarction, revascularization, and cardiac death in patients who present with symptoms suggestive of acute coronary syndrome (ACS) without a diagnosis of myocardial infarction. All specimens were collected under a separate specimen collection protocol (Protocol No, 7B5-02-09A01-01: Clinical Specimen Procurement for Biomarker Evaluation of Suspected ACS). The specimens that will be tested include approximately 8300 specimens collected from 1101 subjects presenting to Emergency Departments with signs and symptoms of Acute Coronary Syndrome and will be provided to the clinical sites performing investigational Troponin I testing.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
1,101
Test blood samples from the ARCHITECT STAT High Sensitive Troponin I Assay. Results obtained will be used to assess the prognosis of subjects for risk of ACM/MACE in the timeframes (30 days and 90 days) after the Emergency Department visit. Troponin results will be compared to documented ACM/MACE events at the 30 day and 90 day time points after Emergency Department visit.
Chandler Regional Medical Center
Chandler, Arizona, United States
Stanford University School of Emergency Medicine
Palo Alto, California, United States
Nationwide Laboratory Services
Fort Lauderdale, Florida, United States
Clinical Performance - Area Under the Curve
The ARCHITECT STAT High Sensitive Troponin-I assay clinical performance was evaluated by calculating the Area Under the Curve (AUC). The Area Under the Curve that was assessed, is used to determine the optimum clinical sensitivity and specificity for the ARCHITECT STAT High Sensitive Troponin-I assay. The ARCHITECT STAT High Sensitive Troponin-I assay results were generated from subject's specimens collected at three collection time points in three tube types (K2 EDTA, Lithium Heparin Separator and Serum Separator).
Time frame: Troponin value from three collection time points (0-2 hours, >2 and up to 4 hours, >4 hours and up to 9 hours) from subject presentation to the emergency department
Clinical Performance- Sensitivity
The ARCHITECT STAT High Sensitive Troponin-I assay clinical performance was evaluated by calculating Sensitivity. The ARCHITECT STAT High Sensitive Troponin-I assay results were generated from subject's specimens collected at three collection time points in three tube types (K2 EDTA, Lithium Heparin Separator and Serum Separator).
Time frame: Troponin value from three collection time points (0-2 hours, >2 and up to 4 hours, >4 hours and up to 9 hours) from subject presentation to the emergency department
Clinical Performance- Specificity
The ARCHITECT STAT High Sensitive Troponin-I assay clinical performance was evaluated by calculating Specificity. The ARCHITECT STAT High Sensitive Troponin-I assay results were generated from subject's specimens collected at three collection time points in three tube types (K2 EDTA, Lithium Heparin Separator and Serum Separator).
Time frame: Troponin value from three collection time points (0-2 hours, >2 and up to 4 hours, >4 hours and up to 9 hours) from subject presentation to the emergency department.
Clinical Performance- Negative Predictive Value (NPV)
The ARCHITECT STAT High Sensitive Troponin-I assay clinical performance was evaluated by calculating Negative Predictive Value. The ARCHITECT STAT High Sensitive Troponin-I assay results were generated from subject's specimens collected at three collection time points in three tube types (K2 EDTA, Lithium Heparin Separator and Serum Separator).
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
John T Mather Memorial Hospital
Port Jefferson, New York, United States
Duke University Medical Center
Durham, North Carolina, United States
Cleveland Clinic Foundation
Cleveland, Ohio, United States
Pennsylvania State University- Hershey Medical Center
Hershey, Pennsylvania, United States
Hospital of the University of Pennsylvania
Philadelphia, Pennsylvania, United States
Hospital Physicians in Clinical Research
Austin, Texas, United States
Hospital Physicians in Clinical Research
Bryan, Texas, United States
...and 2 more locations
Time frame: Troponin value from three collection time points (0-2 hours, >2 and up to 4 hours, >4 hours and up to 9 hours) from subject presentation to the emergency department.
Clinical Performance- Positive Predictive Value (PPV)
The ARCHITECT STAT High Sensitive Troponin-I assay clinical performance was evaluated by calculating the Positive Predictive Value. The ARCHITECT STAT High Sensitive Troponin-I assay results were generated from subject's specimens collected at three collection time points in three tube types (K2 EDTA, Lithium Heparin Separator and Serum Separator).
Time frame: Troponin value from three collection time points (0-2 hours, >2 and up to 4 hours, >4 hours and up to 9 hours) from subject presentation to the emergency department.
Prognosis
Specimens were collected at 11 EDs from 1,101 subjects presenting to the ED with symptoms consistent with ACS. All subject diagnoses were adjudicated by three board certified cardiologists according to current standard of care. ARCHITECT STAT High Sensitive Troponin I results were generated from subject specimens and evaluated for use as an aid in the assessment of prognosis. Analyses used the first available troponin result from multiple serial draw time points. Subjects were assessed for risk of all cause mortality (ACM) and major adverse cardiac event (MACE) at 30 day and 90 day time points after ED discharge. MACE consisted of myocardial infarction, urgent revascularization, and cardiac death. Subjects were followed-up for subsequent events by medical record review and/or subject/caregiver contact. Any ACM and MACE that occurred during the ED visit and on the same day as ED discharge were not included in the analyses.
Time frame: 30-day and 90-day follow-up