The primary objective of the present trial is to compare the radiation attenuation provided by XPF caps (0.5mm lead-equivalent and 0.3mm lead-equivalent) to the standard protection (fabric cap, basically no protection) as measured in % radiation attenuation) during 150 days of fluoroscopy guided cardiology interventions. The second objective is to compare the operator comfort (rated on a visual analog scale) of wearing the XPF protection caps compared to the standard fabric caps.
The primary objective of the present trial is to compare the radiation attenuation provided by XPF caps (0.5mm lead-equivalent and 0.3mm lead-equivalent) to the standard protection (fabric cap, basically no protection) as measured in % radiation attenuation) during 150 days of fluoroscopy guided cardiology interventions. The second objective is to compare the operator comfort (rated on a visual analog scale) of wearing the XPF protection caps compared to the standard fabric caps. For more information please contact Dr. Raul Herrera, MD, Director Baptist Cardiac and Vascular Institute Baptist Hospital of Miami.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
OTHER
Masking
DOUBLE
Enrollment
197
Interventionalist wear a XPF 0.5mm lead equivalent cap during fluoroscopy guided interventions. XPF is a trademark of BloXR (Salt lake City, UT) for a new bi-layer (Barium/Bismuth) radiation attenuation material.
Interventionalist wear a XPF 0.3mm lead equivalent cap during fluoroscopy guided interventions. XPF is a trademark of BloXR (Salt lake City, UT) for a new bi-layer (Barium/Bismuth) radiation attenuation material.
Interventionalist wear a standard fabric cap during fluoroscopy guided interventions.
Baptist Hospital of Miami, Miami Cardiac and Vascular Institute
Miami, Florida, United States
Radiation Attenuation in Percentage (%)
The null hypothesis is that the XPF 0.5mm lead-equivalent caps attenuation is equal to the XPF 0.3mm lead-equivalent and standard cap attenuations. The alternative hypothesis is that the XPF 0.5mm lead-equivalent cap attenuation is superior or inferior to the XPF 0.3mm lead-equivalent and standard cap attenuations. The groups will be compared separately and a pooled analysis, comparing XPF 0.5mm and 0.3mm combined vs. standard is also planned.The radiation protection (radiation dose reduction) expressed as a percentage was calculated by subtracting radiation measured underneath the cap from radiation measured outside the cap and then dividing this difference by the product of radiation measured outside the cap multiplied by 100.
Time frame: 1 day
Absolute Radiation Exposure Outside of Cap
The cumulative radiation doses were defined as the summation of all correspondent equivalent doses measured.
Time frame: 1 day
Operator Comfort Assessment
After completion of the last procedure on each day, the operators were asked to rate the comfort of wearing the cap on a scale from 0 (unbearably heavy, badly fitting) to 100 (very light, well fitting).
Time frame: 1 day per participant, a total of 197 operator days
Operator Comfort Assessment Based on Weight
After completion of the last procedure on each day, the caps worn by the operators were weighed to assess comfort of wearing the standard cap, XPF 0.3mm and XPF 0.5mm caps. The operators were asked to rate the comfort of wearing the cap on a scale from 0 (unbearably heavy, badly fitting) to 100 (very light, well fitting).
Time frame: 1 day per participant, a total of 197 operator days