The purpose of the study is to show equivalent performance of the Revaclear dialyzer when compared to a dialyzer with larger membrane surface area. Study design: open, randomized, cross-over, multicentric, controlled prospective Medical devices: Revaclear 200 versus FX 60 or Revaclear 400 versus FX 100, depending on patient needs Patients/sample size: 30 adult chronic hemodialysis patients Treatment: Each patient will be treated by hemodialysis for one week (3 dialysis sessions) with Revaclear dialyzers and one week (3 dialysis sessions) with FX dialyzers. Objectives: intraindividual comparison of dialysis dose; reduction rates and total removal of urea, phosphate, creatinine and ß2-microglobulin; albumin loss Primary variable: dialysis dose Kt/V urea Secondary variable: reduction rates and total removal of urea, phosphate, creatinine and ß2-microglobulin Safety variable: albumin loss, blood count
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
30
Each patient is treated by hemodialysis for one week (3 sessions) with each dialyzer type (Revaclear or FX). The order of dialyzers used will be randomly assigned to the patient at randomization.
Dialyseinstitut Prim. Dr. W. Gießauf GmbH
Graz, Austria
Medical University Graz
Graz, Austria
Dialysis dose Kt/V urea
Calculation from pre and post plasma urea concentration or urea nitrogen, session length, ultrafiltration volume and post dialysis weight. Dialysis dose does not have a unit of measure.
Time frame: 2 weeks (6 consecutive dialysis sessions)
Reduction rate of urea, creatinine, phosphate and ß2-microglobulin
Calculation from pre and post dialysis plasma concentrations. Unit of measure is percentage (%).
Time frame: 2 weeks (6 consecutive dialysis sessions)
Total removal of creatinine, phosphate and ß2-microglobulin
Calculation from measured concentration in integrated dialysate sample over entire treatment, dialysate flow rate, treatment duration and ultrafiltration volume. Unit of measure is g or mg per treatment.
Time frame: 2 weeks (6 consecutive dialysis sessions)
Albumin loss
Calculation from measured concentration in integrated dialysate sample over entire treatment, dialysate flow rate, treatment duration and ultrafiltration volume. Unit of measure is g or mg per treatment.
Time frame: 2 weeks (6 consecutive dialysis sessions)
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.