The ABSORB III RCT is a prospective randomized, single-blind, multi-center trial. It is the pivotal trial to support the US pre-market approval (PMA) of Absorb™ Bioresorbable Vascular Scaffold (BVS). The ABSORB III includes additional two trials i.e. ABSORB III PK (pharmacokinetics) sub-study and ABSORB IV RCT trial which are maintained under one protocol because both trial designs are related, ABSORB IV is the continuation of ABSORB III and the data from ABSORB III and ABSORB IV will be pooled to support the ABSORB IV primary endpoint. Both the trials will evaluate the safety and effectiveness of Absorb BVS.
ABSORB III RCT: A. Primary Objective: The pivotal trial to support the US pre-market approval (PMA) of Absorb BVS. ABSORB III will evaluate the safety and effectiveness of the Absorb BVS System compared to the XIENCE in the treatment of subjects, including those with diabetes mellitus, with ischemic heart disease caused by up to two de novo native coronary artery lesions in separate epicardial vessels. B. Powered Secondary Objectives: 1. Lead-In Phase Objective: To evaluate the applicability and transferability of the didactic Absorb BVS physician training plan to US clinical practice. The lead-in phase is a non-randomized, single-arm, open label group of up to 50 subjects treated with Absorb BVS at up to 35 US sites. The Lead-In phase will enroll/register subjects prior to the randomization phase of ABSORB III. The Lead-In Phase allows the treatment of up to two de novo native coronary artery lesions in different epicardial vessels with reference vessel diameter (RVD) ≥ 2.75 mm to ≤ 3.25 mm and lesion lengths ≥ 8 to ≤ 14 mm. 2. Imaging Cohort Objective: To evaluate long-term vascular function and patency of the Absorb BVS treated segments compared to XIENCE treated segments in the treatment of subjects with ischemic heart disease caused by up to two de novo native coronary artery lesions in separate epicardial vessels. The imaging cohort-phase is a prospective, randomized (2:1 Absorb BVS to XIENCE), single-blind, multi-center trial, registering approximately 200 subjects. This includes 150 subjects for the angiographic/intravascular ultrasound (IVUS) endpoints analysis and approximately 50 subjects for optical coherence tomography (OCT) endpoints analysis. The 200 subjects are separate from the 2000 subjects included in the primary analysis. Data from two powered secondary endpoints from this cohort will support label claims of superiority of Absorb BVS as compared to XIENCE specific to vasomotion and late lumen enlargement. All other subjects in ABSORB III unless specified will receive treatment of up to two de novo native coronary artery lesions in different epicardial vessels with RVD ≥ 2.5 mm to ≤ 3.75 mm and lesion lengths ≤ 24 mm.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
2,008
* Scaffold diameters: 2.5, 3.0 and 3.5 mm * Scaffold lengths: 8, 12, 18, and 28 mm The 3.0 x 18 mm Absorb BVS will be used for the Lead-In. Both the 8 mm and 12 mm lengths will be available for the 2.5/3.0 mm diameter Absorb BVS. Only the 12 mm length will be available for the 3.5 mm diameter. * The commercially approved CE marked device will be used in geographies where it is commercially available. The commercially approved CE marked 23mm Absorb BVS device will not be used in this study. Bioabsorbable drug eluting stent implantation for improving coronary luminal diameter in patients, including those with diabetes mellitus, with ischemic heart disease due to de novo native coronary artery lesions (length ≤ 24 mm) with a reference vessel diameter of ≥ 2.5 mm and ≤ 3.75 mm.
Commercially approved XIENCE Family Stent System, inclusive of XIENCE V, XIENCE PRIME, XIENCE Xpedition, XIENCE Alpine, XIENCE Pro (OUS only), and XIENCE ProX (OUS only). * Stent diameters: 2.5, 2.75, 3.0, 3.25, 3.5 and 4.0 mm * Stent lengths: 8, 12, 15, 18, 23, and 28 mm. The 3.25 mm is only available for XIENCE Xpedition * For geographies where these devices are commercially available, the investigational sties may use only their locally approved devices To improve coronary luminal diameter in patients, including those with diabetes mellitus, with ischemic heart disease due to de novo native coronary artery lesions (length ≤ 24 mm) with a reference vessel diameter of ≥ 2.5 mm and ≤ 3.75 mm.
Baptist Medical Center Princeton
Birmingham, Alabama, United States
University of Alabama Hospital
Birmingham, Alabama, United States
Thomas Hospital
Fairhope, Alabama, United States
Baptist Medical Center South
Montgomery, Alabama, United States
Chandler Regional Medical Center
Gilbert, Arizona, United States
Number of Cardiac Death/TV-MI/ID-TLR (TLF)
TLF is defined as composite of Cardiac Death, Myocardial Infarction (per protocol-defined MI definition), attributable to Target Vessel (TV-MI), or Ischemic-Driven Target Lesion Revascularization (ID-TLR).
Time frame: 1 year
Number of Participants With Powered Secondary Endpoint: Angina
Angina is defined as the first adverse event resulting in the site diagnosis of angina.
Time frame: 1 year
Number of Participants With Powered Secondary Endpoint: All Revascularization
This powered secondary endpoint is intended to assess all revascularization at 1 year and test for superiority of Absorb BVS to XIENCE. All revascularizations are comprised of TLR, TVR excluding TLR, and non-TVR.
Time frame: 1 year
Number of Participants With Powered Secondary Endpoint: Ischemia Driven Target Vessel Revascularization (ID-TVR)
This powered secondary endpoint is intended to assess all ID-TVR at 1 year and test for superiority of Absorb BVS to XIENCE.
Time frame: 1 year
Acute Success- Device Success (Lesion Level Analysis)
Successful delivery and deployment of the study scaffold/stent at the intended target lesion and successful withdrawal of the delivery system with attainment of final in-scaffold/stent residual stenosis of less than 30% by quantitative coronary angiography (QCA) (by visual estimation if QCA unavailable). When bailout scaffold/stent is used, the success or failure of the bailout scaffold/stent delivery and deployment is not one of the criteria for device success.
Time frame: On day 0 (the day of procedure)
Acute Success: Procedural Success (Subject Level Analysis)
Achievement of final in-scaffold/stent residual stenosis of less than 30% by QCA (by visual estimation if QCA unavailable) with successful delivery and deployment of at least one study scaffold/stent at the intended target lesion and successful withdrawal of the delivery system for all target lesions without the occurrence of cardiac death, target vessel MI or repeat TLR during the hospital stay (maximum of 7 days).
Time frame: On day 0 (the day of procedure)
Number of Death (Cardiac, Vascular, Non-cardiovascular)
DEATH (Per ARC Circulation) : All deaths are considered cardiac unless an unequivocal non-cardiac cause can be established. Specifically, any unexpected death even in patients with coexisting potentially fatal non-cardiac disease (e.g. cancer, infection) should be classified as cardiac. Cardiac death (CD): Any death due to proximate cardiac cause (e.g. MI, low-output failure, fatal arrhythmia), unwitnessed death and death of unknown cause, all procedure related deaths including those related to concomitant treatment. Vascular death: Death due to non-coronary vascular causes such as cerebrovascular disease, pulmonary embolism, ruptured aortic aneurysm, dissecting aneurysm, or other vascular cause. Non-cardiovascular death: Any death not covered by the above definitions such as death caused by infection, malignancy, sepsis, pulmonary causes, accident, suicide or trauma.
Time frame: 0 to 5 years
Number of Participants With All Myocardial Infarction (MI)
* Attributable to target vessel (TV-MI) * Not attributable to target vessel (NTV-MI)
Time frame: 0 to 5 years
Number of Participants With All Target Lesion Revascularization (TLR)
TLR is defined as any repeat percutaneous intervention of the target lesion or bypass surgery of the target vessel performed for restenosis or other complication of the target lesion. All TLR should be classified prospectively as ischemia driven (ID-TLR) or not ischemia driven (NID-TLR) by the investigator prior to repeat angiography. The target lesion is defined as the treated segment from 5 mm proximal to the stent and to 5 mm distal to the stent.
Time frame: 0 to 5 years
Number of Participants With All Target Vessel Revascularization (TVR) Excluding Target Lesion Revascularization (TLR)
TVR is defined as any repeat percutaneous intervention or surgical bypass of any segment of the target vessel. The target vessel is defined as the entire major coronary vessel proximal and distal to the target lesion which includes upstream and downstream branches and the target lesion itself.
Time frame: 0 to 5 years
Number of Participants With All Revascularization
All revascularization endpoint is comprised of TLR, TVR excluding TLR, and non-TVR.
Time frame: 0 to 5 years
Number of Death/All MI
All deaths includes Cardiac death: Any death due to proximate cardiac cause (e.g. MI, low-output failure, fatal arrhythmia), unwitnessed death and death of unknown cause, all procedure related deaths including those related to concomitant treatment. Vascular death: Death due to non-coronary vascular causes such as cerebrovascular disease, pulmonary embolism, ruptured aortic aneurysm, dissecting aneurysm, or other vascular cause. Non-cardiovascular death: Any death not covered by the above definitions such as death caused by infection, malignancy, sepsis, pulmonary causes, accident, suicide or trauma. Myocardial Infarction (MI) - Q wave MI: Development of new, pathological Q wave on the ECG. -Non-Q wave MI: Those MIs which are not Q-wave MI
Time frame: 0 to 5 years
Number of Cardiac Death/All MI
All deaths includes Cardiac death: Any death due to proximate cardiac cause (e.g. MI, low-output failure, fatal arrhythmia), unwitnessed death and death of unknown cause, all procedure related deaths including those related to concomitant treatment. Vascular death: Death due to non-coronary vascular causes such as cerebrovascular disease, pulmonary embolism, ruptured aortic aneurysm, dissecting aneurysm, or other vascular cause. Non-cardiovascular death: Any death not covered by the above definitions such as death caused by infection, malignancy, sepsis, pulmonary causes, accident, suicide or trauma. Myocardial Infarction (MI) - Q wave MI: Development of new, pathological Q wave on the ECG. -Non-Q wave MI: Those MIs which are not Q-wave MI
Time frame: 0 to 5 years
Number of Cardiac Death/TV-MI/ID-TLR (TLF)
Target Lesion Failure is composite of Cardiac death/ Target Vessel Myocardial Infarction (TV-MI)/ Ischemic-Driven Target Lesion Revascularization (ID-TLR).
Time frame: 0 to 5 years
Number of Cardiac Death/All MI/ID-TLR (Major Adverse Cardiac Events-MACE)
Major adverse cardiac events (MACE) is defined as the composite of cardiac death, all myocardial infarction, and ischemic driven target lesion revascularization (ID-TLR).
Time frame: 0 to 5 years
Number of Participants With Target Vessel Failure (TVF)
Target Vessel Failure (TVF) is the composite of Cardiac Death, Myocardial infarction (MI) or Ischemic-Driven Target Vessel Revascularization (ID-TVR).
Time frame: 0 to 5 years
Number of Death/All MI/All Revascularization (DMR)
DMR is the composite of All Death, All Myocardial infarction (MI) and All Revascularization.
Time frame: 0 to 5 years
Number of Participants With Acute Stent/Scaffold Thrombosis (Per ARC Definition)
Stent Thrombosis should be reported as a cumulative value at the different time points and with the different separate time points. Time 0 is defined as the time point after the guiding catheter has been removed and the subject left the Catheterization lab. Timing : Acute : 0 - 24 hours post stent implantation; Subacute : \>24 hours - 30 days post stent implantation; Late : 30 days - 1 year post stent implantation; Very late : \>1 year post stent implantation. Evidence: Definite stent thrombosis is considered to have occurred by either angiographic or pathologic confirmation. Probable stent thrombosis is considered to have occurred after intracoronary stenting in case of Any unexplained death within the first 30 days or Irrespective of the time after the index procedure, any MI that is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of stent thrombosis and in the absence of any other obvious cause.
Time frame: ≤ 1 Day
Number of Participants With Acute/Subacute Stent/Scaffold Thrombosis (Per ARC Definition)
Stent Thrombosis should be reported as a cumulative value at the different time points and with the different separate time points. Time 0 is defined as the time point after the guiding catheter has been removed and the subject left the Catheterization lab. Timing : Acute : 0 - 24 hours post stent implantation; Subacute : \>24 hours - 30 days post stent implantation; Late : 30 days - 1 year post stent implantation; Very late : \>1 year post stent implantation. Evidence: Definite stent thrombosis is considered to have occurred by either angiographic or pathologic confirmation. Probable stent thrombosis is considered to have occurred after intracoronary stenting in case of Any unexplained death within the first 30 days or Irrespective of the time after the index procedure, any MI that is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of stent thrombosis and in the absence of any other obvious cause.
Time frame: 0 to 30 Days
Number of Participants With Subacute Stent/Scaffold Thrombosis
Stent Thrombosis should be reported as a cumulative value at the different time points and with the different separate time points. Time 0 is defined as the time point after the guiding catheter has been removed and the subject left the Catheterization lab. Timing : Acute : 0 - 24 hours post stent implantation; Subacute : \>24 hours - 30 days post stent implantation; Late : 30 days - 1 year post stent implantation; Very late : \>1 year post stent implantation. Evidence: Definite stent thrombosis is considered to have occurred by either angiographic or pathologic confirmation. Probable stent thrombosis is considered to have occurred after intracoronary stenting in case of Any unexplained death within the first 30 days or Irrespective of the time after the index procedure, any MI that is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of stent thrombosis and in the absence of any other obvious cause.
Time frame: >1 to 30 Days
Number of Participants With Late Stent/Scaffold Thrombosis (Per ARC Definition)
Stent Thrombosis should be reported as a cumulative value at the different time points and with the different separate time points. Time 0 is defined as the time point after the guiding catheter has been removed and the subject left the Catheterization lab. Timing : Acute : 0 - 24 hours post stent implantation; Subacute : \>24 hours - 30 days post stent implantation; Late : 30 days - 1 year post stent implantation; Very late : \>1 year post stent implantation. Evidence: Definite stent thrombosis is considered to have occurred by either angiographic or pathologic confirmation. Probable stent thrombosis is considered to have occurred after intracoronary stenting in case of * Any unexplained death within the first 30 days or * Irrespective of the time after the index procedure, any MI that is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of stent thrombosis and in the absence of any other obvious cause.
Time frame: 31 to 365 Days
Number of Participants With Very Late Stent /Scaffold Thrombosis (Per ARC Definition)
Stent Thrombosis should be reported as a cumulative value at the different time points and with the different separate time points. Time 0 is defined as the time point after the guiding catheter has been removed and the subject left the Catheterization lab. Timing : Acute : 0 - 24 hours post stent implantation; Subacute : \>24 hours - 30 days post stent implantation; Late : 30 days - 1 year post stent implantation; Very late : \>1 year post stent implantation. Evidence: Definite stent thrombosis is considered to have occurred by either angiographic or pathologic confirmation. Probable stent thrombosis is considered to have occurred after intracoronary stenting in case of * Any unexplained death within the first 30 days or * Irrespective of the time after the index procedure, any MI that is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of stent thrombosis and in the absence of any other obvious cause.
Time frame: 366 to 393 Days
Number of Participants With Cumulative Stent/Scaffold Thrombosis
Stent Thrombosis should be reported as a cumulative value at the different time points and with the different separate time points. Time 0 is defined as the time point after the guiding catheter has been removed and the subject left the Catheterization lab. Timing : Acute : 0 - 24 hours post stent implantation; Subacute : \>24 hours - 30 days post stent implantation; Late : 30 days - 1 year post stent implantation; Very late : \>1 year post stent implantation. Evidence: Definite stent thrombosis is considered to have occurred by either angiographic or pathologic confirmation. Probable stent thrombosis is considered to have occurred after intracoronary stenting in case of * Any unexplained death within the first 30 days or * Irrespective of the time after the index procedure, any MI that is related to documented acute ischemia in the territory of the implanted stent without angiographic confirmation of stent thrombosis and in the absence of any other obvious cause.
Time frame: 0 to 1853 Days
Pre-Procedure Minimum Lumen Diameter (MLD)
Angiographic endpoint Minimum lumen diameter is defined as the shortest diameter through the center point of the lumen.
Time frame: < or = 1 day
Pre-Procedure Percent Diameter Stenosis (%DS)
Percent Diameter Stenosis is defined as the value calculated as 100 \* (1 - Minimum Luminal Diameter (MLD)/Reference vessel diameter (RVD)) using the mean values from two orthogonal views (when possible) by quantitative coronary angiography (QCA).
Time frame: < or = 1 day
Post-Procedure In-Segment Minimum Lumen Diameter (MLD)
Angiographic endpoint. Minimum lumen diameter is defined as the shortest diameter through the center point of the lumen. In- Segment is defined as, within the margins of the stent or scaffold and 5 mm proximal and 5 mm distal to the stent or scaffold.
Time frame: ≤ 7 days post index procedure
Post-Procedure In-Segment Percent Diameter Stenosis (%DS)
Angiographic endpoint. Percent Diameter Stenosis is defined as the value calculated as 100 \* (1 - Minimum Luminal Diameter (MLD)/Reference vessel diameter (RVD)) using the mean values from two orthogonal views (when possible) by quantitative coronary angiography (QCA). In- Segment is defined as, within the margins of the stent or scaffold and 5 mm proximal and 5 mm distal to the stent or scaffold.
Time frame: ≤ 7 days post index procedure
Post-Procedure In-Device Minimum Lumen Diameter (MLD)
Angiographic endpoint. Minimum lumen diameter is defined as the shortest diameter through the center point of the lumen. In- Segment is defined as, within the margins of the stent or scaffold and 5 mm proximal and 5 mm distal to the stent or scaffold
Time frame: ≤ 7 days post index procedure
Post-Procedure In-Device Percent Diameter Stenosis (%DS)
Angiographic endpoint. Percent Diameter Stenosis is defined as the value calculated as 100 \* (1 - Minimum Luminal Diameter (MLD)/Reference vessel diameter (RVD)) using the mean values from two orthogonal views (when possible) by quantitative coronary angiography (QCA).
Time frame: ≤ 7 days post index procedure
Post-Procedure In-Device Acute Gain
The acute gain was defined as the difference between post- and pre procedural minimal lumen diameter (MLD).
Time frame: ≤ 7 days post index procedure
Powered Imaging Cohort Secondary Endpoint: The Instent/Scaffold Mean Lumen Area Change, From Post Procedure to 3 Years by Intravascular Ultrasound (IVUS)
* Mean lumen area measured after nitrate infusions, superiority test, \~300 pooled subjects. * Pooled IVUS subjects (\~300 subjects): 150 subjects from the Imaging Cohort of ABSORB III RCT and 150 subjects from ABSORB Japan RCT.
Time frame: From Post procedure to 3 Years
Optical Coherence Tomography (OCT) Endpoint: Mean Neointimal Area (NIA)
All OCT endpoints will be collected for within the device and within the treated segment: Descriptive analysis of strut, lesion and vessel morphology Mean neointimal area (NIA) - Apposed to the vessel wall with neointimal coverage Apposed to vessel wall without neointimal coverage Incomplete apposition to vessel wall with neointimal coverage Incomplete apposition to vessel wall without neointimal coverage Lumen area/volume stenosis % Mean/minimal device area Mean/minimal luminal area/volume Mean strut area/volume Persisting incomplete apposition, late incomplete apposition at 3 years (if analyzable) OCT analysis for subjects with jailed side branch Descriptive analyses from 3-dimensional OCT reconstructions
Time frame: 3 Years
Optical Coherence Tomography (OCT) Endpoint: Mean Device Area, Adluminal
All OCT endpoints will be collected for within the device and within the treated segment: Descriptive analysis of strut, lesion and vessel morphology Mean neointimal area (NIA) - Apposed to the vessel wall with neointimal coverage Apposed to vessel wall without neointimal coverage Incomplete apposition to vessel wall with neointimal coverage Incomplete apposition to vessel wall without neointimal coverage Lumen area/volume stenosis % Mean/minimal device area Mean/minimal luminal area/volume Mean strut area/volume Persisting incomplete apposition, late incomplete apposition at 3 years (if analyzable) OCT analysis for subjects with jailed side branch Descriptive analyses from 3-dimensional OCT reconstructions
Time frame: 3 Years
Optical Coherence Tomography (OCT) Endpoint: Mean Lumen Area
All OCT endpoints will be collected for within the device and within the treated segment: Descriptive analysis of strut, lesion and vessel morphology Mean neointimal area (NIA) - Apposed to the vessel wall with neointimal coverage Apposed to vessel wall without neointimal coverage Incomplete apposition to vessel wall with neointimal coverage Incomplete apposition to vessel wall without neointimal coverage Lumen area/volume stenosis % Mean/minimal device area Mean/minimal luminal area/volume Mean strut area/volume Persisting incomplete apposition, late incomplete apposition at 3 years (if analyzable) OCT analysis for subjects with jailed side branch Descriptive analyses from 3-dimensional OCT reconstructions
Time frame: 3 Years
Optical Coherence Tomography (OCT) Endpoint: Minimal Lumen Area
All OCT endpoints will be collected for within the device and within the treated segment: Descriptive analysis of strut, lesion and vessel morphology Mean neointimal area (NIA) - Apposed to the vessel wall with neointimal coverage Apposed to vessel wall without neointimal coverage Incomplete apposition to vessel wall with neointimal coverage Incomplete apposition to vessel wall without neointimal coverage Lumen area/volume stenosis % Mean/minimal device area Mean/minimal luminal area/volume Mean strut area/volume Persisting incomplete apposition, late incomplete apposition at 3 years (if analyzable) OCT analysis for subjects with jailed side branch Descriptive analyses from 3-dimensional OCT reconstructions
Time frame: 3 Years
Optical Coherence Tomography (OCT) Endpoint: Percentage of Malapposition Struts
All OCT endpoints will be collected for within the device and within the treated segment: Descriptive analysis of strut, lesion and vessel morphology Mean neointimal area (NIA) - Apposed to the vessel wall with neointimal coverage Apposed to vessel wall without neointimal coverage Incomplete apposition to vessel wall with neointimal coverage Incomplete apposition to vessel wall without neointimal coverage Lumen area/volume stenosis % Mean/minimal device area Mean/minimal luminal area/volume Mean strut area/volume Persisting incomplete apposition, late incomplete apposition at 3 years (if analyzable) OCT analysis for subjects with jailed side branch Descriptive analyses from 3-dimensional OCT reconstructions
Time frame: 3 Years
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Banner Heart Hospital
Mesa, Arizona, United States
Banner Good Samaritan Medical Center
Phoenix, Arizona, United States
Scottsdale Healthcare
Scottsdale, Arizona, United States
Arkansas Heart Hospital
Little Rock, Arkansas, United States
John Muir Medical Center - Concord Campus
Concord, California, United States
...and 182 more locations