The purpose of this study is to compare multidetector computed tomography (CT) image quality following peripheral venous high pressure injection of iodinated intravenous contrast media between a 20 gauge (GA) fenestrated, intravenous catheter (BD Nexiva™ Diffusics™) and an 18 GA non-fenestrated, intravenous catheter. In the contrast enhanced CT procedure (CECT), a machine (power injector) is used to rapidly deliver contrast agent to an individual's circulatory system. The contrast allows for greater differentiation of tissues in images of the area of interest, affording a better opportunity to visually assess the area and hence make a more accurate diagnosis. CECT often requires a high rate of contrast flow (≥ 5 mL/sec) through the IV catheter to provide sufficient contrast for an acceptable image. Typically an 18 GA IV catheter is required to achieve these high flow rates; however, often the patient does not have a vein that will accommodate this large size catheter. The BD Nexiva™ Diffusics™ IV catheter is designed to deliver a higher flow rate than a conventional catheter of the same gauge by incorporating fenestrations (small openings) near the tip of the catheter that allow fluid to exit the catheter in addition to that which normally exits through the main opening at the tip of the catheter.
This is a prospective study performed in adult human subjects who have been referred for a CECT. The study will occur during a single visit. Once consented and enrolled, subjects whose antecubital veins can accommodate an 18 GA IV catheter (as assessed by the Inserter or designee of the PI), will be randomized to receive either a conventional 18 GA IV catheter (reference, control) or a 20 GA BD Nexiva™ Diffusics™ IV catheter (test). Subjects whose veins are not considered suitable for an 18 GA IV catheter will be placed in a non-randomized cohort for placement of a 20 GA BD Nexiva™ Diffusics™ IV catheter. Subjects will receive one (1) catheter for the study. The primary objective of this study is to compare multidetector computed tomography (CT) image quality following peripheral venous high pressure injection of iodinated intravenous contrast media between a 20 GA fenestrated, intravenous catheter (BD Nexiva™ Diffusics™) and an 18 GA non-fenestrated, intravenous catheter. The 20 GA BD Nexiva™ Diffusics™ IV catheter will be considered non-inferior to conventional 18 GA IV catheters if the 95% upper bound for the difference in percentage of images of acceptable quality between the images acquired using the 18 GA IV catheters and the 20 GA BD Nexiva™ Diffusics™ IV catheters is less than 15%. Study catheters will be inserted by appropriately licensed healthcare professionals (Inserters). Prior to enrollment, potential Inserters will be trained and qualified by the Sponsor on proper use of the BD Nexiva™ Diffusics™ IV catheter in both an artificial hand model and then in human patients. Inserters will then be expected to use the test product for a period of two (2) weeks, in order to gain familiarity with the product. Because BD Nexiva™ Diffusics™ IV catheters are not currently used in the Johns Hopkins Outpatient Center, the Sponsor will provide BD Nexiva™ Diffusics™ IV catheters for use during this training and familiarization period.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
238
The Inserter will have a single attempt on each arm (antecubital area) in which to insert the study catheter. If the catheter is not successfully inserted on the first attempt, the Inserter may make a second attempt in the contralateral arm. If the Inserter is unsuccessful in the second attempt, the subject is to be discontinued from the Study, and additional (non-Study) insertion attempts may be made according to facility policies.
The Inserter will have a single attempt on each arm (antecubital area) in which to insert the study catheter. If the catheter is not successfully inserted on the first attempt, the Inserter may make a second attempt in the contralateral arm. If after the first attempt, there is no vein in the contralateral arm that can accommodate an 18 gauge catheter, the subject is to be discontinued from the Study, and additional (non-Study) insertion attempts may be made according to facility policies.
Johns Hopkins Outpatient Center
Baltimore, Maryland, United States
Acceptable Image Quality
Study images were assessed by a US board-certified radiologist to determine whether the image is of acceptable quality. The radiologist was not informed of the study device used for the injection that produced the image under evaluation. Subjective image quality assessment for acceptability was determined by: * The report of the reading radiologist in the section of the report where the radiologist indicates if the image is acceptable or not. The absence of a comment in this section will be interpreted as acceptable, and, * the assessment of an independent single second reader (such as the subinvestigator or research radiologist) blinded to the reading radiologist's report and the infusion catheter type. In the case of a non-concurrence, the Principal Investigator assessed the image and his or her assessment had final authority.
Time frame: at the time of image assessment
Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Randomized Subjects
Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified.
Time frame: at the time of image assessment
Chest CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects)
Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified.
Time frame: at the time of image assessment
Complete Chest and Abdomen CT - Average Hounsfield Units as a Measure of Aortic Contrast Delivery and Enhancement in Subjects Whose Veins Could Accommodate an 18 GA IV Catheter (Randomized Subjects)
Objective image quality assessment was performed by a board-certified Research Radiologist, by measuring the post-contrast aortic attenuation in Hounsfield Units (HU) at either the aortic arch if chest is imaged or the diaphragmatic crus and above the aortic bifurcation or on the most inferior image on the arterial phase. Hounsfield units within the proximal and distal aorta were recorded for each subject image reviewed. The average number of HU for each group in the Study are reported descriptively; formal acceptance criteria for this objective are not specified.
Time frame: at the time of image assessment
Maximum Flow Rate
Maximum flow rate guidelines provided in the BD Nexiva Diffusics Instructions for Use were to be followed. The maximum flow rate (milliliters per second, or mL/sec) utilized was recorded by the Radiology Technician immediately following the power injection of iodinated intravenous contrast media. Results are descriptively summarized; no formal acceptance criteria were specified by the protocol.
Time frame: immediately after power injection
Catheter Insertion Success
Insertion is successful when the catheter can be flushed or infused to demonstrate patency, and there is no inadvertent administration of a solution or medication into the tissue surrounding the IV catheter. The number of participants with successfully inserted catheters after the first or second insertion attempt is reported. The clinician inserting the IV catheter made a clinical judgement as to whether the catheter was successfully placed; the protocol did not define more specific criteria. Catheter insertion success rates were determined from each IV insertion attempted in the Study.
Time frame: immediately after catheter insertion
Extravasation of Contrast Media
The number of subjects who experienced injections with extravasation of contrast media is reported.
Time frame: upon contrast injection
Automatic Injection Shutoff
The number of subjects who experienced injections with automatic injection shutoff is reported. Immediately after the power injection, the injection technician recorded whether or not an automatic injection shutoff occurred.
Time frame: immediately after contrast injection
High Pressure Alarm
The number of subjects who experienced injections with activation of the high pressure alarm is reported. Immediately following the power injection, the clinician performing the power injection recorded whether or not the high pressure alarm sounded.
Time frame: immediately after contrast injection
Catheter Dislodgement
The number of subjects who experienced partial or complete dislodgement of the catheter from the subject prior to power injection procedure completion is reported. The information about the IV catheter was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection.
Time frame: Immediately following power injection
Catheter Transfixation
The number of subjects who experienced catheter transfixation (the IV penetrating the opposite wall of the vein) is reported. This information was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection.
Time frame: immediately after power injection
Catheter Integrity Failure
Catheter integrity failure generally refers to any portion of the device breaking or malfunctioning so that its proper function is no longer assured. In this study, the most relevant catheter integrity failures to be assessed were fluid leakage, tubing rupture, or tubing separation from the hub. The number of subjects who experienced injections with a catheter integrity failure during injection is reported. The information about the IV catheter was collected by a clinically qualified study staff member (e.g., the Principal or sub-Investigator, or Study Coordinator) after the power injection.
Time frame: immediately after power injection