The purpose of this study was to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
113
Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
Consumer Product Testing Co.
Fairfield, New Jersey, United States
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<75mg/dL) and within +/- 20% (for reference BG results \>=75mg/dL) of the YSI reference method results.
Time frame: 1 hour
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 100 mg/dL) and Within ±15% (≥ 100 mg/dL) of the Laboratory Glucose Method
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<100mg/dL) and within +/- 15% (for reference BG results \>=100mg/dL) of the YSI reference method results.
Time frame: 1 hour
Percent of Self Test Fingerstick Blood Glucose Results Within ±12.5 mg/dL (< 100 mg/dL) and Within ±12.5% (≥ 100 mg/dL) of the Laboratory Glucose Method
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 12.5mg/dL (for reference BG results \<100mg/dL) and within +/- 12.5% (for reference BG results \>=100mg/dL) of the YSI reference method results.
Time frame: 1 hour
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Percent of Blood Glucose Results From Alternative Site Testing (AST) of the Palm Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Method.
Untrained subjects with diabetes self-tested Alternative Site (AST) Palm blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS AST results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) and within +/- 15% (\>=75mg/dL YSI capillary plasma).
Time frame: 1 hour
Percent of Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method When Tested by Study Staff
Study Staff tested subject fingerstick blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<75mg/dL) and within +/- 20% (for reference BG results \>=75mg/dL) of the YSI reference method results.
Time frame: one hour
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree.'
Time frame: 1 hour