The Sensor Optimization of Cardiac Resynchronization Therapy (CRT) Response (SOCR) Study is a multicenter, prospective, non-randomized acute feasibility study that is being conducted to determine if subcutaneous heart sounds and/or intracardiac impedance can acutely identify the optimal atrioventricular (AV) pacing intervals and optimal left ventricular (LV) electrodes in patients indicated for cardiac resynchronization therapy.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
50
All study patients were evaluated for the optimal atrial-ventricular (AV) programmed interval using various methods.
Iowa Heart Center
Des Moines, Iowa, United States
The Ohio State University
Columbus, Ohio, United States
Mercy Hospital Fairfield
Fairfield, Ohio, United States
Victoria Cardiac Arrhythmia Trials
Victoria, British Columbia, Canada
AV Interval Determination Using Impedance
Current practices use the measurement LV dP/dt max to determine how the AV interval should be programmed in a CRT device. Intracardiac impedance is another method that could be used to determine the optimal AV interval. This outcome measure is the number of patients where the optimal AV interval setting as determined by intracardiac impedance agrees within one AV interval setting (30 milliseconds) of the optimal setting determined by LV dP/dt max.
Time frame: During implant
AV Interval Determination Using Heart Sounds
Current practices use the measurement LV dP/dt max to determine how the AV interval should be programmed in a CRT device. A heart sounds measure, called S1 Amplitude Transition, is another method that could be used to determine the optimal AV interval. This outcome measure is the number of patients where the optimal AV interval setting as determined by heart sounds agrees within one AV interval setting (30 milliseconds) of the optimal setting determined by LV dP/dt max
Time frame: During implant
Optimal Electrode Configuration Determination Using Impedance
Current practices use the measurement LV dP/dt max to determine how the optimal electrode configuration for a CRT device. Intracardiac impedance is another method that could be used to determine the optimal electrode configuration. This outcome measure is the number of patients where the optimal electrode configuration setting as determined by intracardiac impedance agrees with the optimal setting determined by LV dP/dt max
Time frame: During implant
Optimal Electrode Configuration Determination Using Heart Sounds
Current practices use the measurement LV dP/dt max to determine how the optimal electrode configuration for a CRT device. Heart sounds, as measured by S1 Amplitude, is another method that could be used to determine the optimal electrode configuration. This outcome measure is the number of patients where the optimal electrode configuration setting as determined by heart sounds agrees with the optimal setting determined by LV dP/dt max
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Ottawa Heart Institute
Ottawa, Ontario, Canada
Institut Universitaire de Cardiologie et de Pneumologie de Québec
Québec, Quebec, Canada
Grantham Hospital
Hong Kong, Hong Kong
Queen Mary Hospital
Hong Kong, Hong Kong
Kings College Hospital
London, United Kingdom
Time frame: During implant