The purpose of this study was to determine if untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
219
Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
AMCR Institute
Escondido, California, United States
Rainier Clinical Research Center
Renton, Washington, United States
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL (<75mg/dL) or Within +/-20% (>=75mg/dL) of Laboratory Glucose Method
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Self-Test Fingerstick Blood Glucose (BG)Results Within +/- 15mg/dL (<100mg/L YSI) or Within +/- 15% (>=100mg/dL YSI) of Laboratory Glucose Method
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI BG results were used to calculate the number of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma)
Time frame: 1 hour
Number of Self-Test Fingerstick Blood Glucose (BG)Results Within +/- 12.5mg/dL (<100mg/L YSI) or Within +/- 12.5% (>=100mg/dL YSI) of Laboratory Glucose Method
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI BG results were used to calculate the number of BGMS results within +/- 12.5mg/dL (\<100mg/dL YSI capillary plasma) or +/- 12.5% (\>=100mg/dL YSI capillary plasma)
Time frame: 1 hour
Number of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/L YSI) or Within +/- 20% (>=75mg/dL YSI) of Laboratory Glucose Method
Untrained subjects with diabetes self-tested Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results were used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma)
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Time frame: 1 hour
Number of Subject Fingerstick BG Results Within +/- 15mg/dL (<75mg/L YSI) or Within +/- 20% (>=75mg/dL YSI) of Laboratory Glucose Method When Tested by Study Staff
Study staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma)
Time frame: 1 hour
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'.
Time frame: 1 hour