The primary purpose of this investigation is to validate the performance and safety of the Investigational Device (ID) against a validated and legally cleared device for measuring Blood Pressure (BP). The secondary purpose is to validate the performance of the ID over a period of 24 hours against a validated and legally cleared device for Ambulatory Blood Pressure Monitoring (ABPM).
The clinical investigation is conducted as a controlled (subject as own control), prospective, randomised, comparative investigation in a population with a representative spread of BPs to ensure that the ID performs as good as current marketed and approved devices. The study will be conducted in compliance with the International Standard ANSI/AAMI/ISO 81060-2 - Non-invasive sphygmomanometers - Clinical validation of automated measurement type. This clinical investigation will enrol up to 96 male and female subjects to ensure the recording of 85/35 evaluable datasets in accordance with the ANSI/AAMI/ISO standard \[11\]. All measurements will be paired and falls into 2 categories: Auscultatory measurements as gold standard reference, oscillometric measurements for clinical practise reference and further validation of ambulatory monitoring and pulse shape measurements.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
SUPPORTIVE_CARE
Masking
NONE
Conventional measuring of BP
New way of measuring BP - ContiPressTM Investigational device
2 different devices for oscillometric measurement of BP
Accuracy of measurement
The reference measurements are compared with the measurements made by the investigational device
Time frame: Up to 4 months
Skin irritation
After removal of each investigational device the skin is assessed. Also participants are instructed to report any skin irritation to investigator
Time frame: Up to 4 months