This study will investigate the effect of wearing a small warming patch, the InsuPatch, at the site of the continuous glucose monitor (CGM), also called a sensor. Specifically, we will be looking to see if using the InsuPatch (IP) improves the accuracy of the continuous glucose monitor.
Hypotheses: 1. Warming of the skin and subcutaneous tissue with the InsuPatch device will improve the accuracy of a CGM as compared to a CGM used alone when both are calibrated with a standard laboratory quality glucose measuring reference standard. 2. Warming of the skin and subcutaneous tissue with the InsuPatch device will reduce the lag time of the CGM, as compared to a CGM used alone when both are calibrated with a standard laboratory quality glucose measuring reference standard. 3. The combination of the InsuPatch device and a CGM will be well-tolerated and not cause skin irritation.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
OTHER
Masking
NONE
Enrollment
11
Device which applies local heating at 40 degrees Celsius to the vicinity of the subcutaneous site of glucose sensor.
Glucose sensor inserted without warming patch
Yale University School of Medicine
New Haven, Connecticut, United States
Point accuracy of the continuous glucose monitor plus Insupatch (CGM+IP) by mean absolute relative difference (MARD) calculation of sensors with and without IP use.
Point accuracy of the continuous glucose monitor plus Insupatch (CGM+IP) by mean absolute relative difference (MARD) calculation of sensors with and without IP use, comparing sensor glucose readings and signal data to serum Yellow Springs Instrument (YSI) glucose measurements, with each subject acting as his or her own control.
Time frame: 4 hours after lunch, 8 hours after breakfast