The BioNIR study aims to show that the BioNIR ridaforolimus eluting stent is non-inferior to the Resolute zotarolimus-eluting stent for the primary clinical endpoint of target lesion failure (TLF) at 12 months; that it is non-inferior to the Resolute for the secondary endpoint of angiographic in-stent late loss at 13 months; and that it is more cost-effective.
The BioNIR is a prospective, multi-center, single-blind, two-arm, randomized clinical trial. The population will consist of subjects undergoing PCI for angina (stable or unstable), silent ischemia, NSTEMI, and recent STEMI. Complex lesions are allowed. There is no limit to the number of lesions per vessel or individual lesion length; however, the total planned stenting in the coronary tree cannot exceed 100mm. Randomization will be stratified by the presence of medically treated diabetes vs. no medically treated diabetes, acute coronary syndrome (ACS) vs. non-ACS, and by site. Lesions planned to be treated must be declared and recorded at time of randomization. Planned staged procedures, if necessary, must be declared immediately post procedure. Clinical follow-up will be performed at 30 days, 6 months, and 1, 2, 3, 4, and 5 years post randomization. 200 patients at participating North American sites will be consented for planned angiographic follow-up at 13 months after enrollment, with 100 of these patients consented to undergo planned IVUS at baseline and at 13 months following randomization. The primary endpoint is Target Lesion Failure (TLF) at 12 months, defined as the composite of cardiac death, target vessel-related myocardial infarction, or ischemia-driven target lesion revascularization. Clinical Secondary Endpoints to be evaluated at 30 days, 6 months, and 1, 2, 3, 4 and 5, except as noted: * Device, Lesion, and Procedure Success at time of baseline procedure * TLF at 30 days, 6 months, and 2, 3, 4 and 5 years defined as the composite of cardiac death, target vessel-related MI, or ischemia-driven TLR. * Major adverse cardiac events (MACE; the composite rate of cardiac death, any MI or ischemia-driven TLR) * Target vessel failure (TVF; the composite rate of death, target vessel related MI or ischemia-driven TVR) * All-cause mortality * Cardiac death * Myocardial Infarction * Target Vessel Related MI * Ischemia-driven TLR * Ischemia-driven TVR * Stent Thrombosis (ARC definite and probable) Angiographic Sub-Study Secondary Endpoint to be evaluated at 13 months: • Angiographic in-stent and in-segment late loss IVUS Sub-Study Secondary Endpoint to be evaluated at 13 months: * In-stent percent neointimal hyperplasia * Stent mal-apposition A key component of this trial will be a prospective assessment of health care resource utilization, costs and cost effectiveness. A separate cost effectiveness assessment plan describes the data collection and analysis. Sample Size Consideration: From recent US trials of best in class DES (Xience V, Promus Element and Resolute), the 1-year TLF rate in patients with non-complex lesions not undergoing routine angiographic follow-up is approximately 3.8%. Using the assumption of the more-comers' design, the 1-year event rate will be conservatively increased by 50% (assuming enrollment rate for complex patients/lesions is 50% with double the standard event rate) - thus 5.8%. Therefore, with a one-sided 95% upper bound of the confidence interval of 3.3% (a relative 57% margin) and 1:1 randomization, enrolling 1810 patients (905 per group) provides 90% power to demonstrate non-inferiority. Assuming 95% follow-up rate at 1 year, approximately 1906 patients will be enrolled (953 in each group).
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
SINGLE
Enrollment
1,919
Piedmont Healthcare
Atlanta, Georgia, United States
ZNA Middelheim
Antwerp, Belgium
Queen Elizabeth II Health Sciences Centre
Halifax, Nova Scotia, Canada
Hadassah Hebrew University Medical Center
Jerusalem, Israel
Target Lesion Failure (TLF)
The primary endpoint of TLF at 12 months was defined as the composite of cardiac death, target vessel-related MI, or ischemia-driven TLR.
Time frame: 12 months
Device Success
Clinical: Acute secondary endpoint determined at time of baseline procedure
Time frame: Determined at time of baseline procedure
TLF
Clinical secondary endpoint to be evaluated at 30 days, 6 months, and 2, 3, 4 and 5 years, defined as the composite of cardiac death, target vessel-related MI, or ischemia-driven TLR
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Major Adverse Cardiac Events
Clinical: MACE; the composite rate of cardiac death, any MI or ischemia-driven TLR
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Target Vessel Failure
Clinical: TVF; the composite rate of death, target vessel related MI or ischemia-driven TVR
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
All Cause Mortality
Clinical: The number of patients who die from all causes
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Cardiac Death
Clinical: The number of patients who die of cardiac-related causes
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Myocardial Infarction
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
San Raffaele Hospital
Milan, Italy
Maasstad Ziekenhuis
Rotterdam, Netherlands
PAKS, II Oddzial Kardiologiczny
Bielsko-Biala, Poland
Hospital Meixoeiro
Pontevedra, VIGO, Spain
Clinical: The number of patients who suffer a myocardial infarction.
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Target Vessel Related MI
Clinical: The number of patients who suffer a MI that is related to the target vessel of the procedure.
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Ischemia Driven TLR
Clinical:
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Ischemia Driven TVR
Clinical:
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Stent Thrombosis
Clinical: ARC definite and probable
Time frame: 30 days, 6 months, and 1, 2, 3, 4 and 5 years
Angiographic Sub-Study: In-stent and In-segment Late Loss
Secondary Endpoint for angiographic in-stent and in-segment late loss
Time frame: 13 months
IVUS Sub-Study: In-stent Percent Neointimal Hyperplasia
IVUS: In-stent percent neointimal hyperplasia
Time frame: 13 months
IVUS Sub-Study: Stent Mal-apposition
IVUS Sub-Study: Stent mal-apposition
Time frame: 13 months
Lesion Success
Measures whether the lesion was successfully treated.
Time frame: Determined at time of baseline procedure
Procedure Success
Acute clinical endpoint: The success of the procedure as determined at time of baseline procedure
Time frame: Determined at time of baseline procedure