The purpose of this study is to determine if subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) without training and obtain valid glucose results.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
SUPPORTIVE_CARE
Masking
NONE
Enrollment
136
Subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi Contour Investigational BG Monitoring System with no training. All BG results are compared to reference method results obtained from subject capillary plasma. Also, study staff test subject venous blood and BG results are compared to reference method results obtained from subject venous plasma.
AMCR Institute
Escondido, California, United States
Rainier Clinical Research Center
Renton, Washington, United States
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Subjects with diabetes self-test fingerstick blood use an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Study staff test subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results are used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).
Time frame: 1 hour
Number of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Subjects with diabetes self-test Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS AST palm results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Study staff obtain and test subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results are used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire Statements
Staff will obtain subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects may respond Strongly Agree; Agree; Neutral; Disagree; or Strongly Disagree.
Time frame: 1 hour