Capnometer has been used to assess the success of intubation and the quality of cardiopulmonary resuscitation. However, capnometers can malfunction under the conditions such as pulmonary edema and hemorrhage due to the vulnerability of capnometers for water. Investigators hypothesized that the use of 10cm catheter mount will reduce the malfunction of capnometers under the conditions of water.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
BASIC_SCIENCE
Masking
NONE
Enrollment
25
Hanyang University
Seoul, South Korea
the malfunction rate of capnograph
The malfunction of capnograph is defined that the ETCO2 level is out-of the measured range of each patient under water-free conditions. The calculation formula of malfunction rate is as follows: malfunctioning time of capnograph / total experiment time (2min). 1. Water conditions in tracheal tube * Water 5ml * Water 10ml 2. Capnometers * Mainstream capnometer * Microstream capnograph 3. Comparison group * No-DCCM group (proximal capnometer of DCCM) * DCCM group (distal capnometer of DCCM) Abbreviation : DCCM, Direct connect catheter mount
Time frame: 1 day
The cumulative survival rate of capnometers under the conditions of water
1\. Kaplan-Meier analysis 1-1. Water conditions in tracheal tube * Water 5ml * Water 10ml 1-2. Capnometers * Mainstream capnometer * Microstream capnograph 1-3. Comparison group * No-DCCM group (proximal capnometer of DCCM) * DCCM group (distal capnometer of DCCM) Abbreviation : DCCM, Direct connect catheter mount
Time frame: 1 day
The consistency of ETCO2 between No-DCCM group and DCCM group under the water-free conditions
1. Bland-Altman plots under water 0ml in tracheal tube. 1-1. Capnometers * Mainstream capnometer * Microstream capnograph 1-3. Comparison group * No-DCCM group (proximal capnometer of DCCM) * DCCM group (distal capnometer of DCCM) 2. The intraclass correlation coefficient (ICC) under water 0ml in tracheal tube 2-1. Capnometers * Mainstream capnometer * Microstream capnograph 2-2. Comparison group * No-DCCM group (proximal capnometer of DCCM) * DCCM group (distal capnometer of DCCM)
Time frame: 1 day