Using an adult lung bench model of non invasive ventilation, the aim of the study is to compare an experimental system of breath-synchronized vibrating mesh nebulizer to a conventional vibrating mesh nebulizer during non invasive ventilation in terms of inhaled and lost doses.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
TREATMENT
Masking
NONE
Enrollment
3
500 mg/4 mL Amikacin nebulized using vibrating mesh nebulizer associated with a single limb bilevel ventilator. The nebulizations are considered as finished when there is no visible evidence of nebulization for a period of 30 seconds. The NIV is set with an IPAP of 15 cmH2O and EPAP of 5 cmH2O. The lung model is simulated with a respiratory rate of 15 cycles/minute
500 mg/4 mL Amikacin nebulized using vibrating mesh nebulizer associated with a single limb bilevel ventilator. The nebulizations are considered as finished when there is no visible evidence of nebulization for a period of 30 seconds. The NIV is set with an IPAP of 15 cmH2O and EPAP of 5 cmH2O. The lung model is simulated with a respiratory rate of 25 cycles/minute
University Hospital St Luc
Brussels, Belgium
Inhaled dose
The inhaled dose assessed by residual gravimetric method
Time frame: after 24 hours
Expiratory wasted dose
The dose expelled in the ambient air through the exhalation port assessed by residual gravimetric method
Time frame: after 24 hours