The objective of this clinical study is to investigate whether the NeuroSENSE is an adequate monitor of hypnotic depth-of-anesthesia (DOA). Therefore, this study will investigate whether the information provided by the NeuroSENSE Monitor can help in assessing the hypnotic effect of anesthetics in adult patients undergoing general anesthesia.
In particular, the study will focus on the correlation between the WAVcns index, a proprietary quantifier of cortical activity displayed by the NeuroSENSE, with (1) commonly assessed clinical endpoints/variables relating to the DOA, (2) changes in anesthetic drug administration, and (3) the anesthesiologist's assessment of the DOA based on standard of care monitors and his/her own observations of the patient state. The hypothesis under evaluation is that the WAVcns index is an independent assessor of the progression of the anesthetic state and events of general anesthesia related to hypnotic endpoints.
Study Type
OBSERVATIONAL
Enrollment
76
Anesthesia in all patients will be induced with iv propofol 1.5 mg/kg with extra 0.5 mg/kg doses as needed. Anesthesia will be maintained using inhaled desflurane in oxygen/air with the initial desflurane concentration of 1.2 MAC, which will be changed in steps of ±0.4 MAC after reaching the post-intubation steady state. The step changes will be initiated every 7.5 min, as follows: at every level of MAC there is a 1/3 chance in remaining at this level and a 2/3 chance of changing it. The anesthesiologist can accept the change to move to the new level. However, he/she can also reject it to remain at the current level for another 7.5 min. This scheme yields 3 possible desflurane levels: 0.8, 1.2 and 1.6 MAC. The anesthesiologist can also leave the randomization schedule at anytime if clinically indicated, and later re-start it at his/her discretion. All anesthetic doses and their changes fall within conventional dosing.
The NeuroSENSE NS-701 system is a 2-channel device for brain activity monitoring in the operating room, intensive care unit, emergency room and other clinical settings. The system acquires and processes electroencephalograms (EEGs) via noninvasive electrodes placed on a patient's forehead. The acquired raw EEG signals and processed EEG variables are continuously displayed for interpretation by the clinician. The proprietary processed variable, WAVcns, quantifies the patient's brain activity, which is typically affected by anesthetic drugs. The system displays processed variables based on 2 bilateral channels (1 per brain hemisphere) for use as a supplement to the anesthesia standard of care. Note: the indications provided by the NeuroSENSE monitor are not used for anesthesia dosing or patient assessment in this study, and the anesthesia provider is blinded to the monitor readings. The device is used only for acquisition of EEG signals and may be used to log the events of interest.
Fraser Health: Royal Columbian Hospital
New Westminster, British Columbia, Canada
Fraser Health: Eagle Ridge Hospital
Port Moody, British Columbia, Canada
Prediction probability Pk for the WAVcns for consciousness vs unconsciousness
The primary outcome will test the performance of WAVcns index in discriminating between consciousness and unconsciousness during induction of anesthesia. It will be evaluated using Receiver-Operating Characteristic (ROC), and Area Under Curve (AUC) will be calculated. Note that the AUC statistic is equal to the prediction probability Pk for dichotomous responses, such as consciousness and unconsciousness.
Time frame: During induction of anesthesia, an expected average of 4 min
Prediction probability Pk for the WAVcns for responders vs nonresponders to lidocaine aerosol
This secondary outcome will test if the WAVcns index prior to lidocaine insufflations or intubation is a good predictor of patient response. The aerosolized lidocaine will be used as a standardized stimulus to objectively assess response observed at the larynx and seen in the patient during airway manipulation during induction of anesthesia.
Time frame: During induction of anesthesia, an expected average of 4 min
Correlation of the WAVcns with primary anesthetic dosing
This secondary outcome will test if the WAVcns index during anesthesia maintenance correlates with primary anesthetic dosing (end tidal agent concentration).
Time frame: During anesthesia maintenance, an expected average of 2 hrs
Prediction probability for the WAVcns for responders vs nonresponders to verbal command
This secondary outcome will test if the WAVcns index accurately indicates return of consciousness (response to verbal command) during emergence.
Time frame: During emergence from anesthesia, an expected average of 20 min
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Non-invasive blood pressure (BP), heart rate (HR), respiratory rate (RR) and electroencephalogram (EEG) will be recorded during the surgery in all subjects. Also, information about all medically significant events, all study-related events (incl. patient's reactions, responses, observations and assessments) and administered medications (incl. end-tidal (ET) desflurane concentration) will be recorded along with time stamps. The anesthesia and physiological parameters (e.g., BP, HR, RR, ET agent concentration) outputted from the anesthesia monitor will be recorded electronically during the surgery by the anesthesia monitoring system (data-points will be recorded at least every 3 min for BP and every 10 sec for other variables). The EEG signal will be continuously recorded by NeuroSENSE NS-701 Monitor (described under a separate intervention). The CRC will also be present in the Operating Room (OR) to record and manage the recording of all the required information.
Anesthesia will be reduced to facilitate rapid recovery at the discretion of the anesthesiologist, about 20 minutes before the end of surgery. The desflurane ET will be stepped down from 0.8 MAC by steps of 0.2 MAC approximately every 5 min during the Emergence.