The purpose of the study is to evaluate the safety and effectiveness of the HW005 System in patients listed for cardiac transplantation with refractory, advanced heart failure at risk of death.
The primary endpoint is success at 180 days which is expressed as a simple proportion and defined as alive on the originally implanted HW005 or transplanted or explanted for recovery. Patients must survive 60 days post-explant for recovery to be considered successful. The primary endpoint is not statistically powered. The six eligible implanted cases will be evaluated as a proportion of success accompanied by a 95% exact confidence interval and presented alongside the ADVANCE results for visual comparison.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
6
The HW005 Pump is in implantable centrifugal pump that was designed to provide flows up to 10 L/min. It is a small device which is both lightweight and simple to use.
Kyushu University Hospital
Fukuoka, Kyushu, Japan
Tohoku University Hospital
Sendai, Tohoku, Japan
Osaka University Hospital
Osaka, Japan
National Cerebral and Cardiovascular Center Hospital
Osaka, Japan
Number of Participants Alive on the Implanted HW005 Ventricular Assist System or Transplanted or Explanted for Recovery at 180 Days
The primary endpoint is success at 180 days which is defined as alive on the originally implanted HW005 Ventricular Assist System or transplanted or explanted for recovery. Patients must survive 60 days post-explant for recovery to be considered successful.
Time frame: 180 days
Number of Participants Alive on the Implanted HW005 Ventricular Assist System at 180 Days
Time frame: 180 days
Incidence of All Serious Adverse Events (SAEs) and Unanticipated Adverse Device Effects (UADEs)
SAEs are defined as events that result in death, are life-threatening, result in permanent disability, require medical treatment to prevent permanent disability or require surgical intervention or hospitalization. UADEs are defined as any serious adverse device effect which by its nature, incidence, severity or outcome has not been identified in the current version of the risk analysis report. Events were categorized using the Japanese registry for Mechanically Assisted Circulatory Support (J-MACS) definitions.
Time frame: Through study completion, an average of 44.5 months
Incidence of All Device Failures and Device Malfunctions.
The device failures and malfunctions are the failures on pumping function due to thrombosis inside/outside of the implanted pump, the mechanical failures of the implanted components, the mechanical failures of the external components and the failures due to the user. Events were categorized using the J-MACS definitions.
Time frame: Through study completion, an average of 44.5 months
Change in Kansas City Cardiomyopathy Questionnaire (KCCQ) Quality of Life Score at 6 and 24 Months
The KCCQ is a self-evaluation questionnaire for patients with heart failure to assess the severity of their symptoms how it may affect their lives. Scores range between 0 (worst) - 100 (best). Values in the table were calculated as the value at Month 6 (180 days) minus Screening and at Month 24 (after FU completion) minus Screening. The positive change in KCCQ score indicates an improvement in patient Quality of Life.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
The University of Tokyo Hospital
Tokyo, Japan
Time frame: Screening, Month 6 (180 days) and Month 24 (after follow-up (FU) completion)
Change in NYHA Patient Classification (Patient Functional Status) at 6 and 24 Months
The New York Heart Association (NYHA) Functional Classification places patients with heart failure in one of four categories based on how much they are limited during physical activity. Class I (best): No limitation of physical activity. Class II: Slight limitation of physical activity. Comfortable at rest. Class III: Marked limitation of physical activity. Comfortable at rest. Class IV (worst): Unable to carry on any physical activity without discomfort. Symptoms of heart failure at rest. A shift from a higher class (e.g. Class IV) to a lower class (e.g. Class II) indicates an improvement in patient health and functional status.
Time frame: Screening, Month 6 (180 days) and Month 24 (after FU completion)
Change in Patient 6-minute Walk Distance (Patient Functional Status) at 6 and 24 Months
For the 6-minute walk test, the patient is asked to walk for 6 minutes. The total distance walked is measured. Values in the table were calculated as the value at Month 6 (180 days) minus Screening and at Month 24 (after FU completion) minus Screening. A positive change in 6-minute walk distance indicates an improvement in health and functional status.
Time frame: Screening, Month 6 (180 days) and Month 24 (after FU completion)