This study evaluates three different neuromuscular monitoring devices (acceleromyography, one- or three-dimensional, and electromyography) with regard to their precision to detect residual paralysis after injection of neuromuscular blocking agents and recurrence of paralysis after administration of reversal agents in a clinical setting.
Neuromuscular blocking agents (NMBAs) are routinely used as standard part of a modern, balanced anesthesia regime. A residual duration of action exceeding the end of surgery, also called residual neuromuscular paralysis, is a common undesired side effect and increases the risk for postoperative pulmonary complications such as aspiration or pneumonia delaying patients' discharge from the post-anesthesia care unit. To limit the incidence of residual paralysis in daily anesthesia care, quantitative neuromuscular monitoring is recommended after injection of NMBAs. If a residual effect of a NMBA is detected at the end of surgery, reversal agents such as cholinesterase inhibitors, e.g. neostigmine or a selective relaxant binding agent, e.g. sugammadex can be administered. However, if reversal agents are not adequately dosed, the risk for a residual neuromuscular blockade re-occurs. Accordingly, neuromuscular monitoring is also useful to control the action of the administered reversal agents. Although several techniques of neuromuscular monitoring are established in clinical practice, electromyography (EMG) and acceleromyography (AMG) are the most common quantitative neuromuscular monitoring devices. Electromyography, the gold standard for detecting residual neuromuscular block, is based on measuring summarized spikes of evoked muscle contractions. Acceleromyography measuring the acceleration of evoked muscle contraction is also commercially available and easy to use. This acceleration, however, can be measured both one-dimensionally and three-dimensionally. This study evaluates the three described neuromuscular monitoring devices with regard to their precision to detect residual paralysis after administration of NMBAs and recurrence of neuromuscular blockade after administration of reversal agents in a clinical setting. We plan to include and randomize a total of 200 patients. The study participants will be recruited from patients scheduled for surgery at the Klinikum rechts der Isar der Technischen Universität München, Munich, Germany. In each patient, acceleromyography (either one- or three-dimensional) will be compared with the calibrated electromyography. The findings will help to indicate which neuromuscular monitoring device is most suitable for detecting residual paralysis and recurrent neuromuscular blockade.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
214
Neuromuscular monitoring
Neuromuscular monitoring
Neuromuscular monitoring
Department of Anesthesiology
Munich, Germany
Klinik für Anaesthesiologie, Klinikum rechts der Isar der Technischen Universität München
Munich, Germany
This study evaluates the precision and agreement of TOF-ratios obtained with acceleromyography and electromyography during residual neuromuscular paralysis
Time frame: during surgery (2 hours)
Incidence of re-occurrence of neuromuscular block after administration of reversal agents measured with acceleromyography and electromyography
Time frame: during surgery (2 hours)