It has been shown that close control of blood glucose levels in the intensive care unit patient has benefits for patient morbidity and mortality rates as well as an impact upon discharge times. GlySure has previously completed an evaluation of its intra-vascular glucose monitoring system in patients in a surgical ICU environment after cardiac surgery. This study seeks to evaluate the same equipment being used in a wider range of patients with a range of underlying conditions who require treatment in medical ICU's. The GlySure device consists of a sensor that is placed into the patient's blood through a central venous catheter. It measures blood glucose levels continuously which allows for more rapid control of blood glucose levels to be achieved as compared to existing methods where blood samples are taken repeatedly every 15 minutes or so. The study is designed to show that the sensor can be used for a protracted period of time, giving accurate results when compared to a reference technique using the i-STAT device. The investigators intend to show that the device performs safely for the required time within the ICU.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
DIAGNOSTIC
Masking
NONE
Enrollment
60
Comparator device is an iSTAT device from Abbott Diagnostics
Safety in Use
No Serious Adverse Events causally related to the device
Time frame: During ICU Stay, normally less than 7 days
MARD Score
Average aggregate MARD to be below 10%, 95% of subjects to have a MARD of less than 25%
Time frame: During ICU stay, normally less than 7 days