The purpose of this study was to determine if untrained persons with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid blood glucose results.
Study Type
INTERVENTIONAL
Allocation
NON_RANDOMIZED
Purpose
SUPPORTIVE_CARE
Masking
NONE
Enrollment
372
Untrained subjects WITH and WITHOUT diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
Untrained subjects WITH diabetes use the Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
AMCR Institute
Escondido, California, United States
Rainier Clinical Research Center
Renton, Washington, United States
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Untrained subjects WITH diabetes (329) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Study staff tested venous blood of subjects WITH diabetes (329) using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).
Time frame: 1 hour
Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff
Study staff obtained and tested subject (329 WITH diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Staff obtained responses from persons WITH diabetes (329) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
Time frame: 1 hour
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15% of the laboratory method across the entire tested YSI glucose range.
Time frame: 1 hour
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-20% of the laboratory method across the entire tested YSI glucose range.
Time frame: 1 hour
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff
Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15% across the tested YSI glucose range.
Time frame: 1 hour
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff
Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-20% across the tested YSI glucose range.
Time frame: 1 hour
Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Staff obtained responses from subjects WITH and WITHOUT diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
Time frame: 1 hour