The overall goal is to evaluate the predictive accuracy of a measure of autoregulatory adequacy derived from CVInsight and compare it with other measures in recognizing hypotensive events during hemodialysis..
Specific Aims are as follows: 1. Assess the accuracy of a new rule set for CVInsight® in predicting intradialytic hypotensive episodes. 2. Assess the comparative predictive accuracy of CVInsight® to other measures with regard to intradialytic hypotensive events. 3. Assess estimated dry weight goals by monitoring post-dialytic vascular refill using CVInsight® and other measures. 4. Assess hemodialysis(HD)-induced myocardial injury by monitoring the occurrence and frequency of arrhythmias and the levels of cardiac Troponin T.
Study Type
OBSERVATIONAL
Enrollment
48
Tufts Medical Center
Boston, Massachusetts, United States
Number of intradialytic hypotensive (IDH) episodes that can be detected by the new rule set for CVInsight.
IDH defined by at least one of the following: 1) Symptoms of hypotension: dizziness, cramping, altered consciousness, nausea +/- vomiting (new since beginning of session), chest pain, shortness of breath, seizure, diaphoretic. 2) Intervention administered due to IDH; relevant interventions:a) a reduction in ultrafiltration (UF) goal programmed at start of treatment (tx); b) admin. of IV fluid, e.g. saline, mannitol or hypertonic saline; c) stopping tx early due to IDH symptoms or event. 3) A significant drop in blood pressure, regardless of symptoms, defined as: * For pts w/ a pre-dialysis systolic blood pressure (SBP) \>110 mm Hg, an intra or post-dialysis SBP \<90 mm Hg, or a drop of \>40 mm Hg over 30 min. * For pts w/ a pre-dialysis SBP \<110, a decline in SBP \<85 mm Hg or 15 mm Hg or lower than their starting BP.
Time frame: 3-4 hours, for two hemodialysis session, or 6-8 hours total
The comparative, predictive accuracy of CVInsight® and CRIT-LINE III with regard to intradialytic hypotensive events.
CVInsight® and CRIT-LINE III will have different predictive accuracies in recognizing IDH episodes. Receiver Operator Characteristics analysis of CVInsight and CRIT-LINE CVInsight: Event level thresholds will be triggered based upon new proprietary rule set. CRIT-LINE: Change in hematocrit of 8% per hour or 15% over the treatment.
Time frame: 3-4 hours, for two hemodialysis session, or 6-8 hours total
Amount of post-dialytic vascular refill as separately indicated by both CVInsight® and CRIT-LINE III.
Post-dialytic vascular refill will indirectly reflect hydration status and accuracy of dry weight estimation. Vascular refill -Ten minutes prior to the end of treatment, with UFR at zero, and BFR at same rate as during treatment, measurements will continue to be taken via the CVInsight® and CRIT-LINE devices to assess vascular refill. CVInsight: Increase in pulse amplitude. CRIT-LINE: Increase of 0.5% increase in hematocrit (HCT) over 10 minutes.
Time frame: At the end of each dialysis session after the ultrafiltration rate (UFR) has turned off.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
The occurrence and frequency of arrhythmias as indicated on the Zio ECG Patch and the levels of cardiac Troponin T, taken prior to each HD treatment. which are indicative HD-induced myocardial injury..
Arrhythmias frequency overall, on dialysis days, during long and short interdialytic periods, nocturnal, diurnal. Troponin T levels will correlate with the incidence rates of arrhythmias. Troponin T as measured by ng/mL, reference value at \<0.011
Time frame: 24 hours before first session of monitored dialysis to at least 24 hours after the second session on monitored dialysis sessions.