The study objectives are to assess the safety and efficacy of the CoreValve Evolut R transcatheter aortic valve replacement (TAVR) system in patients with severe symptomatic aortic stenosis are considered at high or extreme risk for surgical aortic valve replacement.
This objective will be accomplished by a prospective, single arm, historical controlled, multi-site study involving a minimum of 150 implanted subjects with no more than 250 implanted subjects at up to 25 study sites in the United States. Procedural and 30 day safety and efficacy results from this study will be compared to appropriate historical control data for the Medtronic CoreValve System. Subjects will be followed up to 5 years following implantation. The enrollment phase of the study is estimated to take approximately 6-9 months. As each implanted subject is to be followed up to 5 years, the estimated study duration is approximately 66-69 months, excluding the time required for preparing the final report.
Study Type
INTERVENTIONAL
Allocation
NA
Purpose
TREATMENT
Masking
NONE
Enrollment
241
Banner Good Samaritan Medical Center
Phoenix, Arizona, United States
All-cause Mortality at 30 Days by Percent
Percentage of patients that died by any cause at 30 days
Time frame: Assessed at 30 days post-implantation
Percentage of Patients With Disabling Stroke at 30 Days
Stroke Diagnostic Criteria: \> Acute episode of focal or global neurological deficit with at least 2 of the following: * change in level of consciousness \> * hemiplegia, hemiparesis * numbness or sensory loss affecting 1 side \> * dysphasia or aphasia * hemianopia * amaurosis fugax \> * other neurological signs or symptoms consistent with stroke 2.) No other readily identifiable non-stroke cause or the clinical presentation, to be determined by or in conjunctions with the designated neurologist 3.) Confirmation of the diagnosis by at least 1 of the following: * Neurological specialist \> * Neuroimaging procedure, or on clincial grounds alone \> Stroke: durations of neural deficit \> 24 h if available neuroimaging documents a new hemofrrhage or infarct; or the neurological deficit results in death Defined by VARC II: \> An mRS (Modified Rankin Score) of 2 or more at 90 days and an increase in at least 1 mRS category from pre-stroke baseline
Time frame: Assessed at 30 days post-implantation
Percent Device Success Rate Between 24 and 7 Day
Percentage of patients with Device Success defined as: * Absence of procedural mortality, AND * Correct positioning of a single Evolut R valve into the proper anatomical location, AND * Absence of patient-prosthesis mismatch, and mean gradient , 20 mm Hg (or peak velocity \< 3m/sec, AND * Absence of moderate or severe prosthetic valve regurgitation
Time frame: Assessed at 24 hours to seven days post implantation
Percentage of Patients With Less Than Moderate Prosthetic Regurgitation at Early Post Procedure Echocardiogram (24 Hours to 7 Days)
Percentage of patience with none, trace or mild total prosthetic regurgitation at early post procedure echo cardiogram (24 hours to 7 days) as evaluated by echo core lab.
Time frame: Assessed at 24 hours to 7 days post implantation
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
University of Southern California
Los Angeles, California, United States
Yale New Haven Hospital
New Haven, Connecticut, United States
Washington Hospital Center/Medstar
Washington D.C., District of Columbia, United States
St. Vincent Heart Center of Indiana/The Care Group, LLC
Indianapolis, Indiana, United States
University of Kansas
Kansas City, Kansas, United States
The Johns Hopkins University
Baltimore, Maryland, United States
Beth Israel Deaconcess Medical Center
Boston, Massachusetts, United States
University of Michigan Health Systems
Ann Arbor, Michigan, United States
Detroit Medical Center Cardiovascular Institute
Detroit, Michigan, United States
...and 13 more locations
Individual Component of VARC II Safety Endpoint: Percentage of People Requiring Valve-related Dysfunction Requiring Repeat Procedure (BAV, TAVI, or SAVR)
Percentage of patients with any valve dysfunction that requires repeat procedure (e.g. balloon valvuloplasty, TAVI, or surgical AVR), per VARC II definition.
Time frame: Assessed at 30 days post-implantation
Coronary Artery Obstruction Requiring Intervention.
Angiographic or echocardiographic evidence of a new, partial or complete, obstruction of a coronary ostium, either by the Evolut R prosthesis itself, the native leaflets, calcifications, or dissection, occurring during or after the TAVI procedure.
Time frame: Assessed at 30 days post-implantation
Percent VARC II Combined Safety Endpoint at 30 Days
VARC II composite safety endpoint rate includes percent freedom from the following components: * All-cause mortality * All stroke (disabling and non-disabling) * Life-threatening bleeding * Acute kidney injury: stage 2 or 3 (including renal replacement therapy). * Coronary artery obstruction requiring intervention. * Major vascular complication. * Valve-related dysfunction requiring repeat procedure (BAV, TAVI, or SAVR)
Time frame: Assessed at 30 days post-implantation
Percent of Patients With Acute Kidney Injury: Stage 2 or 3 (Including Renal Replacement Therapy).
Stage 2 1. Increase in serum creatinine to 200%-299% (2.0%-2.99% increase compared with baseline) OR \> 2. Urine output \<0.5 mL/kg/h for \>12 but \<24 h \> Stage 3 \> 1\) Increase in serum creatinine to ≥300% (\>3 x increase compared with baseline) OR serum creatinine of ≥4.0 mg/dL (≥354 mmol/L) with an acute increase of at least 0.5 mg/dL (44 mmol/L) OR * Urine output \<0.3 ml/kg/h for ≥24 h OR \> * Anuria for ≥12 h
Time frame: Assessed at 30 days post-implantation
Percentage of Patients With Life-threatening or Disabling Bleeding Event Rate
1. Fatal bleeding (BARC type 5) OR \> 2. Bleeding in a critical organs, such as intracranial, intraspinal, \> intraocular, or pericardial necessitating pericardiocentesis, or intramuscular with compartment syndrome (BARC type 3b and 3c) OR \> 3. Bleeding causing hypovolemic shock or severe hypotension requiring vasopressors or surgery (BARC type 3b) OR \> 4. Overt source of bleeding with drop in hemoglobin ≥5 g/dL or whole blood or packed red blood cells (RBCs) transfusion ≥4 units\* \> (BARC type 3b)
Time frame: Assessed at 30 days post-implantation
Percent Rate of Patients Who Received New Permanent Pacemaker Implant at 30 Days
Percent of patients who underwent implantation of new permanent pacemaker or ICD during or after index procedure
Time frame: Assessed at 30 days
Percent Resheath and Recapture Success Rate
Resheath or recapture success rate (where attempted) where a successful resheath is defined as the intended portion of the Evolut R is resheathed into the capsule of the delivery catheter to the intended amount, as verified by flouroscopy; and a successful recapture is defined as the entire Evolut R TAV (including the frame) is full resheathed into the capsule of the delivery catheter until there is no gap between the capsule and the tip , as verified by flouroscopy. Resheath or recapture wa attempted in a subset of patients. Success rate is calculated as successful resheath or recaputure events in the number of total events. Resheath anad recapture is only possible during the index procedure.
Time frame: Assessed intra-procedurally
Hemodynamic Performance -Mean Gradient
Mean gradient by Doppler echocardiography.
Time frame: Assessed at baseline, 30 days, 6 months, and 1 year
Major Vascular Complication
1. Any aortic dissection, aortic rupture, annulus rupture, left ventricle perforation, or new apical aneurysm/pseudoaneurysm OR \> 2. Access-related vascular injury (dissection, stenosis, perforation, rupture, arterio-venous fistula, pseudoaneurysm, hematoma, irreversible nerve injury, compartment syndrome, percutaneous closure device failure) leading to death, life-threatening or major bleeding, visceral ischemia, or neurological impairment OR \> 3. Distal embolization (noncerebral) from a vascular source requiring surgery or resulting in amputation or irreversible end-organ damage OR \> 4. The use of unplanned endovascular or surgery associated with death, major bleeding, visceral ischemia or neurological impairment OR \> 5. Any new ipsilateral lower extremity ischemia documented by patient symptoms, physical exam, and/or decreased or absent blood flow on lower extremity angiogram OR \> 6. Surgery for access site nerve injury OR \> 7. Permanent access related nerve injury
Time frame: Assessed at 30 days post-implantation
Hemodynamic Performance - Aortic Valve Area
Hemodynamic performance by Doppler echocardiography - Aortic Valve Area cm2
Time frame: Assessed at baseline, 30 days, 6 months, and 1 year
Hemodynamic Performance: Total Prosthetic Valve Regurgitation Graded as Moderate or Severe
Hemodynamic performance: the percent of patients who have a degree of total prosthetic valve regurgitation that is moderate or severe.
Time frame: Assessed at 30 days, 6 months, and 1 year