The purpose of the study is to evaluate different ways to program around extracardiac stimulation.
Study Type
OBSERVATIONAL
Enrollment
749
Number of Vectors With Phrenic Nerve Stimulation (PNS)
The primary objective of this study was to characterize the performance of Boston Scientific (BSC) unique pacing vectors vs. BSC/St Jude Medical (STJ) common pacing vectors to prevent phrenic nerve stimulation (PNS) based on the selected pacing cathode. BSC unique vectors are those that are only available with Boston Scientific's (BSC) quadripolar (X4) CRT-Ds while common vectors are those that are available for both BSC and STJ systems.
Time frame: Pre-discharge (up to 7 days post implant)
Final Pacing Treshold at Pre-discharge
Characterize the electrical performance of BSC pacing vectors using pacing thresholds. Summary statistics for LV Tip1 to LV Ring 2 pacing thresholds for the BSC pacing vectors were reported. The electrical measurements were taken using the implanted pulse generator (PG) per standard practice.
Time frame: Pre-discharge (up to 7 days post implant)
Final LV Lead Impedance at Pre-discharge
Characterize the electrical performance of BSC pacing vectors using impedance measurements. Summary Statistics Were Provided for LV Tip 1 to LV Ring 2 Measurements. The electrical measurements were taken using the implanted pulse generator (PG) per standard practice.
Time frame: Pre-discharge (up to 7 days post implant)
Final Sensing Amplitude at Pre-discharge
Characterize the electrical performance of BSC pacing vectors using sensing amplitudes. Summary statistics were provided for LV Tip 1 to LV Ring 2 measurements. The electrical measurements were taken using the implanted pulse generator (PG) per standard practice.
Time frame: Pre-discharge (up to 7 days post implant)
Delivery System Comparison Between ACUITY Pro and the Other Delivery Systems
Characterize the performance of the BSC ACUITY Pro delivery system versus competitive delivery systems in the placement of LV leads. Data collected on the use of the ACUITY Pro delivery system and competitive systems were characterized by presenting summary statistics by system (all competitive devices maybe combined for purposes of reporting), and differences between the ACUITY Pro system and competitive systems were reported as appropriate.
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.
Mercy Gilbert Medical Center
Gilbert, Arizona, United States
Phoenix Cardiovascular Research Group/ Heart & Vascular Center of Arizona
Phoenix, Arizona, United States
St. Joseph's Hospital
Phoenix, Arizona, United States
Yuma Regional Medical Center
Yuma, Arizona, United States
Cardiology Associates of Northeast Arkansas
Jonesboro, Arkansas, United States
Central Arkansas Veterans Healthcare System
Little Rock, Arkansas, United States
Comprehensive Cardiovascular Medical Group
Bakersfield, California, United States
Cardiac Device Rhythm Specialists, Inc
Los Angeles, California, United States
Desert Heart Rhythm
Palm Springs, California, United States
Bradenton Cardiology Center
Bradenton, Florida, United States
...and 50 more locations
Time frame: At time of implantation (0 to 30 days post consent signature)