The purpose of the study is to test the accuracy benefit of having two glucose sensors (over one sensor alone) when they are positioned: 2mm, 10mm, 20mm, or 30mm apart. It is not yet known how close two sensors can be and still work correctly.
An artificial pancreas system will likely require multiple glucose sensing elements to function safely. It is not yet known how close two glucose sensors can be placed and still work properly. Meaning, if one is reading inaccurately,and the other is positioned adjacent to it, will it read inaccurately also, defeating the purpose of having two sensors. Twenty adult subjects with Type 1 diabetes will wear four Medtronic REAL-Time glucose sensors during two separate 10-hour studies. The inter-sensor distances of each pair to be tested will be: 2, 10, 20, and 30mm apart. The sensors will be inserted the day prior to the study day to allow for signal stabilization and calibration. During the study day, arterialized venous blood will be drawn every 15 minutes to measure blood glucose. The sensor values and interstitial values will be recorded from each sensor receiver at each time point. Subjects will be fed two standardized meals and given pre-meal insulin based on their typical insulin to carbohydrate ratio.
Study Type
OBSERVATIONAL
Enrollment
20
Oregon Health & Science University
Portland, Oregon, United States
Difference between very large sensor accuracy errors (≥50% from reference glucose) with the simultaneous use of four sensors vs one sensor.
Time frame: 9-hour collection period, X 2
Difference between mean absolute relative difference with the simultaneous use of four sensors vs one sensor.
Time frame: 9-hour glucose level collection, X 2
This platform is for informational purposes only and does not constitute medical advice. Always consult a qualified healthcare professional.