The purpose of this study is to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes.
This study is a multi-center, randomized, in-clinic trial that aims to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes for at least one year. Subjects are randomized into either Group A or Group B. The groups assigned will determine when the subject will be participating in the in-clinic YSI frequent sample testing (FST). For example on Day 1: Group A begins FST 30 minutes after Enlite 3 Sensor Insertion. Group B begins FST 14 hours after Enlite 3 Sensor Insertion.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Masking
NONE
Enrollment
118
Use of Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen \& arm used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years who have had a diagnosis of type 1 or type 2 diabetes for at least one year.
AMCR Institute
Escondido, California, United States
Diablo Clinical Research
Walnut Creek, California, United States
Barbara Davis Center for Childhood Diabetes
Aurora, Colorado, United States
Barbara Davis Center
Aurora, Colorado, United States
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD)
Enlite 3 Sensor accuracy using two real time devices: 1) 640G Pump and 2) Guardian Mobile with the minimum calibration requirements (every 12 hours after the second calibration) will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 Days
Retrospective Re-Analysis (MARD With 1 Additional Calibration)
Retrospective re-analysis to simulate 640G Pump and Guardian Mobile 1-minute raw data collected by GST3C Transmitters and GST4C Transmitter: Enlite 3 Sensor accuracy with 3-4 calibrations throughout the day (derived from re-analysis of Enlite 3 Sensor data using actual fingerstick values). Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 Days
Retrospective Analysis (MARD for the GSR With Minimum and 1 Additional Calibration)
Retrospective analysis using one GSR: minimum and 3-4 calibrations will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values during YSI frequent sample testing. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 Days