This study was aimed to to evaluate the efficacy of the new multibending ultra-slim endoscope compared with a conventional ultra-slim endoscope for free-hand insertion of an endoscope into the bile duct for direct POC.
Currently available peroral cholangioscopy (POC) is a duodenoscopy-assisted procedure that does not involve directly inserting an endoscope into the biliary tree. A prototype multibending (MB) ultra-slim endoscope has been developed as a dedicated cholangioscope to overcome the technical difficulties of direct POC. In this study, the investigators evaluated the efficacy of the new MB ultra-slim endoscope compared with a conventional ultra-slim endoscope for free-hand insertion of an endoscope into the bile duct for direct POC without the assistance of accessories. The primary outcome was the technical success of free-hand insertion of the endoscope during direct POC, which was defined as successful insertion of the endoscope through the ampulla of Vater and advancement of the endoscope up to the bifurcation or to the obstructed segment of the biliary tree without any accessories within 15 min.
Study Type
INTERVENTIONAL
Allocation
RANDOMIZED
Purpose
DIAGNOSTIC
Masking
SINGLE
Enrollment
92
The multibending (MB) or conventional ultra-slim endoscope was inserted through the mouth. After facing the papillary orifice, the endoscope was inserted directly, without any accessories, into the distal bile duct. The second bending of the MB ultra-slim endoscope was kept in an upward angled position to achieve a more acute angle of the endoscope tip for entering the distal CBD. Then, the endoscope was pushed by steering the first bend into an upward angle ("α" shape) or the endoscope was pulled by steering the first bend into an upward angle and torqueing it counter-clockwise ("u" shape) for insertion into the distal CBD. If the distal bile duct was visualized on an endoscopic view, carbon dioxide insufflation was stopped. The endoscope was advanced into the hilum or the obstructed segment of the extrahepatic bile duct under fluoroscopic and endoscopic control. Then, the diagnostic and therapeutic interventions were performed during direct POC as necessary.
Soonchunhyang University Bucheon Hospital
Bucheon-si, South Korea
The technical success rate of free-hand insertion of an ultra-slim endoscope during direct POC
successful insertion of the endoscope through the AOV and advancement up to the bifurcation or obstructed segment of the biliary tree, without any accessories
Time frame: within 15 min after insertion of endoscope into the mouth
Procedure time required for free-hand insertion
Procedure time required for free-hand insertion
Time frame: the time of oral intubation with the ultra-slim endoscope to the time of reaching the obstructed segment of the biliary tree or bifurcation, within 15 min after insertion of endoscope into the mouth
The technical success of diagnostic and therapeutic interventions
The technical success of diagnostic and therapeutic interventions
Time frame: Within 24 hours after Completion of DPOC
Adverse events related to direct POC
perforation, cholangitis, pancreatitis, and air embolism
Time frame: Within 7 days after DPOC