This study compares the results of current standard coagulation measurement devices to the Quantra System, a new device, using small amount of extra blood obtained during routine blood draws in neurosurgery patients.
HemoSonics is developing a novel POC diagnostic device, the Quantra System, to perform whole blood coagulation analysis and which is suitable for use in surgical and intensive care settings. The Quantra System employs a patented technology named sonorheometry, which was invented at the University of Virginia. Sonorheometry uses ultrasound pulses to characterize the dynamic changes in viscoelastic properties of a blood sample during coagulation and clot lysis. The initial assay performed on the Quantra System will be the surgical assay for monitoring hemostasis during major surgical procedures in adult patients. A clinical study will be conducted to evaluate the analytical performance as well as compared to existing coagulation monitoring technology
Study Type
OBSERVATIONAL
Enrollment
51
Blood specimen collection
University of Virginia
Charlottesville, Virginia, United States
assay inter-instrument variability
Demonstrate assay inter-instrument variability below 15%
Time frame: 1 day
Establish normal ranges
Establish normal ranges
Time frame: 1 day